Gas Chromatography Method for Quantitation of Residual Solvent Impurities in Nanoformulations.
Krishna Kattel, Rachael M Crist, Jeffrey D Clogston
Nanotechnology Characterization Laboratory, Cancer Research Technology Program, Frederick National Laboratory for Cancer Research, Frederick, MD 21701, USA.
Key points
From the paper's conclusion- Notably, the method was found to be suitable for analysis of formic acid-a common solvent in certain nanoformulations and one in which there is no prior gas chromatography method available which does not require this additional sample manipulation-down to approximately 75 ppm.
Abstract
The development and validation of a sensitive, rapid, and specific gas chromatography method for the evaluation of 19 common Class 2 and Class 3 solvents frequently used in nanomedicine formulation is described. Method validation was performed using PerkinElmer's headspace gas chromatograph system with flame ionization detection and an Elite 624 Crossbond 6% cyanopropylphenyl-94% dimethylpolysiloxane or DB-Fatwax-Ultra Inert column with helium as the carrier gas. Validation characteristics such as linearity, spike recovery, method precision, specificity, sensitivity, limit of detection/quantitation, and analyte stability were evaluated. The validated methods showed excellent linearity, with a correlation coefficient > 0.99, and good precision, with intra-day precision < 7.4% for all tested analytes. The percent recoveries ranged 83-104% within the method's quantitation range. In comparison to previously reported methods, the current method has a much shorter equilibration time, higher sensitivity, better separation for many solvents, and a wide concentration detection range. The current method is also perfectly suitable to analyze short chain fatty acids such as formic acid, acetic acid, butyric acid, and valeric acid without requiring additional extraction or derivatization steps. Notably, the method was found to be suitable for analysis of formic acid-a common solvent in certain nanoformulations and one in which there is no prior gas chromatography method available which does not require this additional sample manipulation-down to approximately 75 ppm. Herein, the method is demonstrated using various nanoformulations, including the commercial Doxil formulation as well as several research nanoformulations, including polymeric, cross-linked polymeric, and dendrimer platforms.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Venturi Easy Ambient Ionization Mass Spectrometry Coupled to Gas Chromatography for High-Throughput Quantitation of Acetic Acid Residues in PharmaceuticalsJournal of the American Society for Mass Spectrometry · 2026
- Screening of residual solvents in cocaine seizures by HS-SPME-GC-MS: Application to multi-class solvent profiling.Journal of forensic sciences · 2026
- High-throughput analysis of target furocoumarins in Citrus essential oils by supercritical fluid chromatography with bio-based solventsGreen Analytical Chemistry · 2026
- Development and validation of the analytical method for determination of residual solvents (2-propanol and n-hexane) in the drug substance Inosine Pranobex.BMC chemistry · 2026
- Rational Selection of Solvents in the Manufacture of Lipid-Based Drug Carriers: A Regulatory Perspective.Biomedicines · 2026
- Rapid Quantification of Hydrazine, Methylhydrazine, and 1,1-Dimethylhydrazine in Environmental Samples by Supercritical Fluid Chromatography-tandem Mass Spectrometry With Pre-column Derivatization by 5-Nitro-2-furaldehyde.Journal of separation science · 2026