Development and Verification of a Suspension-Based TXRF Method for Chromium Determination in Feed and Fecal Samples Containing Chromic Oxide as an External Digestibility Marker
Christina Tzvetkova, Lidia Ivanova, Albena Detcheva, Antonina Kovacheva, Miroslav Simeonov, Irina Karadjova
Bulgarian Academy of Sciences · Trakia University
Key points
From the paper's conclusion- In the present study, a rapid method based on total reflection X-ray fluorescence (TXRF) was developed to determine chromium in feed and fecal samples, containing chromic oxide.
Abstract
In the present study, a rapid method based on total reflection X-ray fluorescence (TXRF) was developed to determine chromium in feed and fecal samples, containing chromic oxide. The method uses suspension sample preparation with gallium as an internal standard and does not require chemical reagents or complete sample digestion. Key parameters affecting performance, such as particle size, sample concentration, and acquisition time, were optimized to ensure stable signals and reliable quantification. The method showed a limit of quantification of 24 µg g−1 for Cr2O3 and good precision (relative standard deviations of 3–7% for both feed and fecal samples), at Cr2O3 concentrations in the range of 20–50 mg kg−1. These performance characteristics meets the requirements for digestibility studies. It requires only small sample quantities with minimal preparation. The developed method is suitable for routine analysis, particularly in studies generating large numbers of samples. The accuracy of the method was confirmed through agreement with results obtained using an independent method based on inductively coupled plasma optical emission spectrometry (ICP-OES) after acid digestion. The scatter plot analysis of the results obtained by both methods showed a linear regression line with a slope of 0.978 and a correlation coefficient (R2) of 0.9559, indicating good agreement. The p-value from the paired t-test performed was greater than 0.05, suggesting that the observed differences between paired measurements are not statistically significant at the 95% confidence level. The Bland–Altman analysis demonstrated negligible systematic deviation between the two methods.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of reagents for suspension-assisted total reflection X-ray fluorescence element analysis in biomedical samplesAnalytical and Bioanalytical Chemistry · 2026
- Magnetite-based SPME-SQT-FAAS for trace chromium determination in riceScientific Reports · 2026
- Advancing Quantitative XRF Analysis of Plant Matrices Through Validation and Extension of Element-Specific Correction Factors by Comparison with INAA Measurements.Methods and protocols · 2026
- Thin pellets, uncertain results? - optimizing pellet thickness for accurate EDXRF determination of trace elements in biological matricesSpectrochimica Acta Part B Atomic Spectroscopy · 2026
- Overcoming Incomplete Digestion of Ti-Doped LiFePO4 Cathode Materials: Microwave-Assisted Sample Preparation and Single-Laboratory Validated ICP-OES Determination of Li, Fe, P, and Ti.Molecules (Basel, Switzerland) · 2026
- Comparison of X-ray Microtomography with Transverse Microradiography for Assessment of Enamel Remineralisation.Journal of dentistry · 2026