Micromachines

Computer-Controlled Balloon Manometry Coupled with Inverse Finite Element Analysis for Assessing Chronic Radiation-Induced Rectal Injury in Rats.

Jing Zhang, Yingxuan Lin, Xin Jin, Mingzhu Xie, Xuejun Ma, Yun Deng, Ben Huang, Hui Zhang,

Department of Radiation Oncology, Fudan University Shanghai Cancer Center, Shanghai 200032, China. · Department of Cardiac Surgery, Zhongshan Hospital, Fudan University, Shanghai 200032, China.

Preclinical

Key points

From the paper's conclusion
  • We integrated computer-controlled balloon manometry, inverse finite element analysis, and paired whole-wall histopathology in a preclinical exploratory workflow.

Abstract

Quantitative assessment of chronic radiation-induced rectal injury (RRI) requires methods that relate tissue remodeling to mechanical function. We integrated computer-controlled balloon manometry, inverse finite element analysis, and paired whole-wall histopathology in a preclinical exploratory workflow. Male Sprague-Dawley rats received 0, 12, 20, or 28 Gy pelvic irradiation and were assessed at 3-4 months. Of 32 enrolled rats, 29 contributed tissue assessments and 26 yielded paired mechanical and histopathological data. Balloon-corrected pressure-volume curves were analyzed with a three-parameter Yeoh model using a common axisymmetric geometry. The recorded coefficient c10 and the post hoc index cmax = max(c10, c20) were associated with injury scores (Spearman ρ = 0.5955 and 0.6675, respectively). Differences from correlations for selected pressure measures were not formally tested. Higher cmax also accompanied edema, fibrosis, and vascular stenosis in exploratory comparisons. Model-archive inspection identified a nearly incompressible formulation and an inconsistency between the radius used in the fitted volume expression and the geometric radius. The outputs have not been recomputed; the coefficients remain provisional descriptors of the archived workflow. The contribution is the integration of controlled loading with a histological reference. Selective testability, unvalidated loading safety, the volume-definition discrepancy, and unestablished parameter identifiability prevent interpretation as validated tissue stiffness or clinical performance.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Computer-Controlled Balloon Manometry Coupled with Inverse Finite Element Analysis for Assessing Chronic Radiation-Induced Rectal Injury in Rats — Science Skim