Metrological Reliability and Long-Term Stability of Digital MEMS Accelerometers: Investigating Calibration Factor Drift, Output Data Rate (ODR) and Signal-to-Noise Ratio (SNR) Consistency Under Accelerated Thermal-Hygrometric Aging.
Antonella Gaspari, Luciano Chiominto, Giulio D'Emilia, Laura Fabbiano, Alessio Facello, Emanuela Natale, Andrea Prato, Alessandro Schiavi
Dipartimento di Meccanica, Matematica e Management, Politecnico di Bari, Via Orabona, 4-70125 Bari, Italy. · Dipartimento di Ingegneria Industriale dell'Informazione e di Economia, Università dell'Aquila, P. le Pontieri, 1-67100 L'Aquila, Italy.
Key points
From the paper's conclusion- Calibration factors were compared for one axis at 10 Hz, together with effective output data rate (ODR) and static signal-to-noise ratio (SNR).
Abstract
This study investigates changes in three metrological attributes of eight nominally identical digital MEMS accelerometers from one production lot. After approximately six years of laboratory storage, six devices underwent three unpowered thermal-hygrometric stress protocols, while two devices provided a natural-aging baseline. Calibration factors were compared for one axis at 10 Hz, together with effective output data rate (ODR) and static signal-to-noise ratio (SNR). Arrhenius, Norris-Landzberg and Peck relationships were examined as conditional exposure-equivalence models rather than validated lifetime predictors. Two thermally cycled devices exhibited calibration-factor increases of 7.9% and 37.6%; the other four stressed devices remained compatible within the reported expanded uncertainties. ODR changes remained below 0.5% for all eight devices. The six stressed devices showed SNR decreases of 0.34-1.43 dB, averaging 0.89 dB, whereas the two baseline decreases were 0.34 and 0.20 dB. Stable ODR therefore did not guarantee stable calibration factor. The contribution is a paired assessment of these selected output attributes, not a complete characterization of the accelerometer or identification of the degrading subsystem. The limited sample size precludes population-level reliability estimates.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Frequency Stability Degradation in Quartz and MEMS Oscillators.Sensors (Basel, Switzerland) · 2026
- Comparability of Archived Aging-Stage and Retest Records from Miniature Temperature-Pressure Sensors: Rank, Concordance, and Response Shape.Sensors (Basel, Switzerland) · 2026
- Reliability Assessment of MEMS Gyroscopes via Dual-Mechanism Synergistic Degradation: A Generalized Linear Model with Physics-Informed Wiener ProcessesSensors · 2026
- Power-Law Degradation and Lifetime Interpretation in Microelectronics ReliabilityApplied Sciences · 2026
- A method to analyze mechanical noise in MEMS capacitive accelerometerEngineering Research Express · 2026
- Pooling interval-monitored step-stress accelerated life tests for reliability prediction: A failure-rate-based meta-analysis for the proportional hazards familyarXiv · 2026