Synthetic-to-Real 5D Pose Estimation for Acoustically Levitated Micro-Calibrators in Lab-on-a-Chip Devices.
Zongxie Chen, Zhijin Li, Yasemin A Geiger, Benedikt K Kaufmann, Dilan Kılıç, Valentina Leone, Stefanie Sudhop, Matthias E Rebhan,
Chair of Media Technology, School of Computation, Information and Technology, Technical University of Munich, Arcisstr. 21, 80333 Munich, Germany. · Heinz-Nixdorf-Chair of Biomedical Electronics, School of Computation, Information and Technology, Technical University of Munich, TranslaTUM, Einsteinstr. 25, 81675 Munich, Germany.
Key points
From the paper's conclusion- We present a synthetic-to-real deep learning framework that addresses this gap by estimating the five-dimensional (5D) pose of a trident-shaped micro-calibrator co-trapped with the cells.
Abstract
Established acoustofluidic lab-on-a-chip devices enable contactless, label-free trapping of cells, but do not provide a quantitative readout of a levitated sample's orientation, restricting their analytical reach. We present a synthetic-to-real deep learning framework that addresses this gap by estimating the five-dimensional (5D) pose of a trident-shaped micro-calibrator co-trapped with the cells. Here, 5D pose denotes the two-dimensional (2D) bounding-box center in the image plane and the calibrator's three-dimensional (3D) orientation. Printed by two-photon polymerization with 10, 15, and 20 μm arms, the calibrator provides the rigid, asymmetric geometry that a translucent, deformable spheroid lacks. The framework cascades detection, foreground segmentation, and rotation regression, with the normalized bounding-box center as a positional auxiliary input, and optionally applies temporal smoothing; it is trained on domain-randomized renderings, eliminating manual 3D pose annotation. On 15,000 synthetic test images, it reached a mean geodesic rotation error of 0.82°. The ablation study showed that the positional input substantially improves rotation estimation and background removal further lowers the error, and identified segmentation, not detection, as the dominant upstream error source. Pseudo-label adaptation raised the expert-assessed correctness rate from 34.71% to 83.62% on 5281 real microscopy frames from which the pseudo-labels were drawn (adaptation performance) and from 2.48% to 29.34% on 242 held-out fluorescence frames with adherent cells (generalization performance). This demonstrates the feasibility of a calibrator-based orientation readout on an established acoustofluidic platform without hardware change.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Autonomous Droplet Navigation via Model-Based Reinforcement Learning: Zero-Shot Transfer and Emergent DynamicsarXiv · 2026
- Autonomous Droplet Navigation via Model-Based Reinforcement LearningarXiv · 2026
- ReLink: Reliable Deep Learning for Microscopic Motion Tracking.Advanced science (Weinheim, Baden-Wurttemberg, Germany) · 2026
- AxonSynth: Domain-Randomized Synthetic Data for Zero-Shot 3D Axon Segmentation in Light-Sheet MicroscopyarXiv · 2026
- Electromagnetic Localization 6-DoF Pose Estimation with TGNet-Guided Initial Value RefinementMathematics · 2026
- From Language to Motion: Task-Conditioned Focal-Stack Trajectory Integration for Microscopic RobotsarXiv · 2026