Abstract
Performance and hemocompatibility are the main focus areas in the development of new hemofilters for continuous kidney replacement therapy. The novel multiFlux hemofilter features a polysulfone membrane with increased hydrophilicity to improve performance stability and hemocompatibility. In the present in vitro study, we compared the multiFlux hemofilter to established hemofilters on the market, i.e., Ultraflux AV (polysulfone-based), Prismaflex HF (polyarylethersulfone-based) and Prismaflex ST (AN69 ST-based). For the characterization of performance, dialysate-side plasma clearance of urea, creatinine and β2-microglobulin (β2M) were investigated over 24 h. Hemocompatibility was characterized in a blood recirculation setup by determining activation of platelets, plasmatic coagulation, the complement system and leukocytes. For urea and creatinine clearance, all hemofilters reached or approached the effluent limit across all time points. For β2M, the multiFlux hemofilter showed significantly higher clearance values at each time point vs. all comparators (p < 0.001) and the most stable performance over 24 h. The multiFlux hemofilter also showed a favorable hemocompatibility profile, especially in terms of platelet activation (vs. Ultraflux AV and Prismaflex HF) and complement activation (vs. all comparators). This in vitro study supports previous findings that increased membrane hydrophilicity improves performance stability and hemocompatibility of synthetic membranes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Pharmacokinetic/pharmacodynamic validation of hemodialyzers as cost-effective alternatives to commercial cartridges in the hollow-fiber infection model.International journal of antimicrobial agents · 2026
- Comparison of polymethyl methacrylate and AN69 surface-treated hemofilters in patients with septic acute kidney injury requiring continuous kidney replacement therapy: impact of baseline kidney function on mortalityFrontiers in Nephrology · 2026
- The Role of Online Hemodiafiltration in Contemporary Kidney CareClinical Journal of the American Society of Nephrology · 2025
- Dialysis Membranes and Hemodialyzers.Contributions to nephrology · 2026
- A scalable microfluidic dialyzer enabling low-flow hemodialysis with clinically relevant clearance.Journal of membrane science · 2026
- A literature-informed pharmacokinetic framework for polymyxin B dosing during continuous renal replacement therapy: integrating renal function and extracorporeal clearance.Frontiers in pharmacology · 2026