Insertion Depth of an Esophageal Stethoscope with a Thermistor: Agreement with Tympanic Temperature and an Acceptable Range Referenced to the Maximal Heart-Sound Position.
Hwa Song Jong, Dong Joon Kim, Ki Tae Jung
Department of Anesthesiology and Pain Medicine, Chosun University Hospital, College of Medicine, Chosun University, Gwangju 61453, Republic of Korea.
Key points
From the paper's conclusion- The MHS position is a bedside landmark rather than a predictor of an individual patient’s depth, and the range derived here requires prospective validation.
Abstract
Background and Objectives: The distal esophagus is a preferred site for intraoperative core-temperature monitoring, but there is no consensus on the insertion depth of an esophageal stethoscope with a thermistor (EST). We aimed to identify the depth of closest agreement with tympanic membrane (TM) temperature and the acceptable range around it. Materials and Methods: Forty adults (ASA I-II, 20-65 years) undergoing elective supine surgery under general anesthesia were studied prospectively (ClinicalTrials.gov NCT05552092). After induction, an 18-Fr EST was inserted orally, and the maximal heart-sound (MHS) depth was recorded by auscultation. From 30 min after induction, it was withdrawn in 2-cm steps every 5 min through ten positions (45 to 27 cm), with paired esophageal and TM measurements at each step. Results: The mean TM-esophageal difference rose from -0.27 °C at 45 cm to +0.77 °C at 27 cm, with the population mean crossing zero at 38.0 cm. At each patient's best-agreement step, selected post hoc as the closest of ten, bias was -0.02 °C (95% limits of agreement -0.14 to +0.11 °C), with all patients within ±0.5 °C. The per-patient interpolated zero-crossing depth lay 2.1 cm deeper than the MHS position (p < 0.001). Exploratory analysis showed that placement from MHS -1 cm to at least +4 cm kept ≥90% of measurements within ±0.5 °C. Only the MHS position was associated with the best-agreement step (r = 0.39, p = 0.014), accounting for 15% of its variance. Conclusions: The EST agreed closely with TM temperature over a wide range of depths centered just distal to the MHS position, and shallow placement degraded accuracy faster than deep placement. The MHS position is a bedside landmark rather than a predictor of an individual patient's depth, and the range derived here requires prospective validation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Esophageal thermometry outperforms bladder and tympanic monitoring during targeted temperature management and shock in the ICU.Journal of critical care · 2026
- Accuracy of Ultrasound-Estimated Epidural Depth in Transverse Median Versus Parasagittal Oblique Planes for Lower Thoracic Epidural Anesthesia: A Prospective Observational Comparative Study.Cureus · 2026
- A Comprehensive Review of Non-Invasive Core Body Temperature Measurement TechniquesSensors · 2026
- The shiny tendon of the sternocleidomastoid muscle: An intraoperative anatomical lighthouse for safe navigation in endoscopic thyroidectomy.Journal of minimal access surgery · 2026
- Feasibility of prior‐informed domain‐based electrical impedance tomography for deep tissue thermometry: Phantom experiments and computed tomography‐derived numerical simulationsMedical Physics · 2026
- Ultrasound innovations in diaphragm assessment: an integrative review of expanding clinical applicationsEuropean Respiratory Review · 2025