Human Osteoblast Behavior on Titanium Implant Surface Coating with Sintered Hydroxyapatite
Andreia Bandeira Luis, Narayan Sahoo, Beatriz Ferreira Fernandes, António Mata, Óscar Carvalho, Joana Faria Marques
Key points
From the paper's conclusion- Considering the inherent bounds of this in vitro study, smooth titanium surfaces exhibited the highest osteoblast viability, whereas the 0.25 mm pattern overall induced a more favorable cellular response compared to the 0.8 mm pattern.
Abstract
Osseointegration is strongly influenced by the physical and chemical characteristics of the implant surfaces, namely topography, roughness profile, and biofunctionalization. The current research assessed the influence of laser surface texturing and hydroxyapatite (HA) biofunctionalization on the viability, morphology, and inflammatory profile of human osteoblasts. Ti-6Al-4V disks were mechanically polished or laser-textured using an Nd:YVO4 laser to generate checkerboard patterns with spatial periodicities of 0.25 or 0.8 mm. Textured surfaces were subsequently coated with nanohydroxyapatite and sintered using either conventional furnace or CO2 laser sintering. The immortalized human osteoblastic line (hFOB 1.19) was seeded onto seven experimental surfaces for up to 14 days. Cell viability was assessed by a resazurin-based assay, while cellular attachment and morphological features were examined using scanning electron microscopy. Additionally, release levels of IL-1β, IL-8, and IL-10 were quantified by ELISA. Smooth titanium exhibited higher osteoblast viability than textured surfaces, whereas the 0.25 mm pattern generally performed better than the 0.8 mm pattern. HA biofunctionalization consistently improved viability within both texture patterns, irrespective of the sintering method, while the sintering technique itself had no significant effect. The 0.8 mm pattern showed lower IL-1β secretion, whereas IL-10 remained largely unchanged. Considering the inherent bounds of this in vitro study, smooth titanium surfaces exhibited the highest osteoblast viability, whereas the 0.25 mm pattern overall induced a more favorable cellular response compared to the 0.8 mm pattern. Hydroxyapatite biofunctionalization improved cell viability, whereas no consistent differences were observed between conventional and laser sintering.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Superhydrophilic Hierarchical Anatase Coating on Sandblasted, Acid-Etched Titanium: In Vitro Apatite Formation and Osteoblast Responses and the Role of Polar Surface Free EnergyJournal of Functional Biomaterials · 2026
- TiO2 Nanotubes Co-Functionalized With Hydroxyapatite, Collagen, and Nisin for Bioactive Titanium Surfaces: A Preliminary In Vitro Study.Journal of biomedical materials research. Part B, Applied biomaterials · 2026
- Enhanced Biological Performance of Fractal Geometry Design on Titanium for Dental Implant Application: An In Vitro Study.ACS biomaterials science & engineering · 2026
- Biofunctionalization of SLM Ti–40Nb alloy through hydroxyapatite-modified plasma electrolytic oxidation coatingApplied Surface Science Advances · 2025
- Influence of Ta2O5 Addition on the Surface Characteristics, In Vitro Bioactivity, and Early Cellular Responses of Single-Layer SLM Titanium-Based Surface LayersJournal of Composites Science · 2026
- Apatite-coated implant surfaces exhibit superior biological, immunological, and mechanical properties compared to sandblasted acid-etched surfacesScientific Reports · 2026