Evaluation of the Performance of the BD Veritor™ System for Detection of Acute Respiratory Viruses: Comparison with RT-PCR in Mali
Rabiatou Sanogo, Ibrehima Guindo, Alhadji Alassane Dicko, Noumou Yakhouba Keïta, Lassina Traoré, Demba Koïta, Boubacar Doumbia, Mahamadou Abdou,
Key points
From the paper's conclusion- The reduced sensitivity observed may partly be explained by low viral loads, prolonged frozen storage of samples before antigen testing, and the complex epidemiological context associated with the co-circulation of respiratory viruses during the COVID-19 pandemic.
Abstract
The COVID-19 pandemic highlighted the need for rapid and accurate diagnosis of respiratory viral infections and reinforced the importance of integrated respiratory virus surveillance systems. Rapid diagnostic tests (RDTs) provide timely results but may show limited sensitivity compared with molecular assays. This study aimed to evaluate the diagnostic performance of the BD Veritor™ System for Rapid Detection of Respiratory Viruses for detecting respiratory syncytial virus (RSV), Influenza A, and Influenza B in Mali, using multiplex real-time RT-PCR as the reference standard. A total of 1202 nasopharyngeal and/or oropharyngeal swab specimens collected between March 2020 and March 2022 from suspected respiratory infection cases were analyzed. Samples were tested using the BD Veritor™ rapid antigen assay and compared with the ViroQ® SARS-CoV-2/FluA/FluB/RSV multiplex RT-PCR assay performed on the LightCycler® 480 II system. Diagnostic performance was assessed using sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and Cohen’s kappa coefficient. The BD Veritor™ assay demonstrated consistently high specificity (>91%) for all viral targets, whereas sensitivity remained low, ranging from 21.05% to 30.43%. Positive predictive values ranged from 24.78% to 46.15%, while negative predictive values remained above 83%. Agreement with RT-PCR was low (κ = 0.14–0.24), indicating limited concordance between the two methods. The reduced sensitivity observed may partly be explained by low viral loads, prolonged frozen storage of samples before antigen testing, and the complex epidemiological context associated with the co-circulation of respiratory viruses during the COVID-19 pandemic. Although the BD Veritor™ assay may provide useful support for rapid clinical orientation and triage in resource-limited settings, negative results should be interpreted cautiously and confirmed by RT-PCR whenever possible.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Design and Validation of Multiplexed Respiratory RT-LAMP assay for the differentiation of SARS-CoV-2, Influenza A and Respiratory Syncytial Virus (RSV) in COVID-19 Pandemic Samples from Western Kenya.medRxiv · 2026
- Diagnostic accuracy of rapid integrated molecular testing for major respiratory viral infections: a systematic review and meta-analysis.Frontiers in medicine · 2026
- Diagnostic performance of antigen-based rapid test kits for six respiratory viruses in children: comparison with a multiplex PCR panel and correlation with viral load.Journal of clinical microbiology · 2026
- Comparative evaluation of Allplex RV master assay and biofire respiratory panel 2.1 plus for detection of respiratory virusesScientific Reports · 2026
- Comprehensive Performance Verification of the Mindray MN2880 System in Detecting Respiratory Virus Nucleic Acids.Journal of clinical laboratory analysis · 2026
- Performance of an antigen test for detection of SARS-CoV-2, influenza, and respiratory syncytial virus—a prospective studyClinical Microbiology and Infection · 2026