Cross-Domain Benchmarking of Focus Measures for Smear-Microscopy Autofocus Under a Consensus-Audited Reference.
Dineth Hewavitharana, Dumith Jayathilaka, Palitha Dassanayake, Ranjith Amarasinghe
Department of Mechanical Engineering, Faculty of Engineering, University of Moratuwa, Katubedda 10400, Sri Lanka.
Key points
From the paper's conclusion- We benchmark 32 handcrafted focus measures on 26,100 z-stacks spanning five smear-microscopy domains at native acquisition dimensions under one frozen protocol.
Abstract
Focus-measure recommendations for smear microscopy are typically established on one specimen type under one reference definition and one scoring rule, leaving their transferability untested. We benchmark 32 handcrafted focus measures on 26,100 z-stacks spanning five smear-microscopy domains at native acquisition dimensions under one frozen protocol. A fixed ten-voter plurality consensus (REF-B) supplies an identical reference target for every candidate, and two further constructions, single-voter exclusion (REF-A) and a non-derivative four-voter consensus (REF-C), quantify how far the recommendation depends on that target. Ten criteria covering consensus localization, curve shape, perturbation response and kernel cost form a declared cross-domain score; clustered bootstrap resampling, alternative aggregation, weight perturbation and controlled image resampling separate sampling variability from sensitivity to evaluation policy. Gradient responses lead under every analysis. Variance of Gradient ranks first under the primary policy in all 1000 clustered bootstrap replicates and in 67.9% of 1000 sampled weight configurations, while Brenner Gradient attains the lowest consensus deviation, 0.072 planes, at the lowest measured kernel cost. Localization ordering is preserved exactly when the reference is rebuilt from non-derivative voters alone (Spearman ρ = 1.000), although the reference plane itself shifts substantially. The benchmark gives a reproducible, auditable basis for selecting focus measures under stated reference, scoring and image-sampling conditions.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Metric-driven Pre-acquisition Diagnostics for Automated Microscopy Ultraviolet Surface Excitation (MUSE)bioRxiv · 2026
- Training-free autofocus via turning-point analysis of a frequency-slope metric.Optics express · 2026
- MicroDeblurNet: high-fidelity restoration of spatially variant defocus in microscopic images for cucumber downy mildewPlant Methods · 2026
- Optical-parameter-conditioned one-shot autofocus for multi-objective-lens microscopy.Optics express · 2026
- Autocorrelation-based beam shape optimization in scanning electron microscopyarXiv · 2026
- A Reproducible Microscopy Acquisition Pipeline and Microscopy-Aware Attention for Classification, Detection, and Segmentation of MicroorganismsMachine Learning and Knowledge Extraction · 2026