Journal of functional morphology and kinesiology

Agreement Between Segmental Bioelectrical Impedance Analysis and Dual-Energy X-Ray Absorptiometry (DXA) for the Magnitude and Direction of Between-Limb Lean Mass Asymmetry in Trail Runners.

Nil Piñol-Granadino, Neus Guinovart, Marta Carrasco-Marginet, Sílvia Puigarnau, Fabrizio Gravina-Cognetti, Quim Rosales, Alfredo Irurtia

Institut Nacional d'Educació Física de Catalunya (INEFC), Av. de l'Estadi 12-22, 08038 Barcelona, Spain. · BCN Body Scan SL, Av. de l'Estadi 12-22, 08038 Barcelona, Spain.

Observational

Key points

From the paper's conclusion
  • Conclusions: Segmental BIA cannot replace DXA for assessing between-limb lean-mass asymmetry in individual trail runners.

Abstract

Background: Between-limb lean-mass asymmetry has been proposed for injury-risk and return-to-sport monitoring. Dual-energy X-ray absorptiometry (DXA), the reference method, requires a fixed installation and ionizing radiation; segmental bioelectrical impedance analysis (BIA) is fast, inexpensive and field-applicable. Agreement for the asymmetry itself, rather than absolute limb values, is unresolved. Methods: In this cross-sectional method-comparison study, 127 trail runners (90 men, 37 women) underwent eight-electrode segmental BIA and DXA in the same session. An asymmetry index, the between-limb difference relative to the limb mean, was computed from raw leg resistance (BIA) and leg lean mass (DXA), and assessed by Bland-Altman analysis, concordance correlation (CCC) and kappa. Results: Mean bias was -0.97 percentage points (pp), but the 95% limits of agreement spanned -6.67 to +4.73 pp, three to four times the observed interquartile range; CCC was 0.414. The same leg was identified in 70.9% of runners (kappa 0.309) against a right-side predominance of 75.6%, and the BIA classification was not significantly better than a constant right-leg rule (p = 0.451). Conclusions: Segmental BIA cannot replace DXA for assessing between-limb lean-mass asymmetry in individual trail runners. The disagreement between methods exceeds the asymmetries observed, and directional agreement must be read against the base rate.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Agreement Between Segmental Bioelectrical Impedance Analysis and Dual-Energy X-Ray Absorptiometry (DXA) for the Magnitude and Direction of Between-Limb Lean Mass Asymmetry in Trail Runners — Science Skim