Ti-Containing Mesh Implants Modified by Heterostructures Based on Noble Metal: Fabrication and Study of Biological Activity.
Svetlana Dorovskikh, Evgeniia Vikulova, David Sergeevichev, Tatiana Guselnikova, Aleksey Kolodin, Anastasiya Fedorenko, Alexander Zheravin, Natalya Morozova
Nikolaev Institute of Inorganic Chemistry, Siberian Branch, Russian Academy of Sciences, Ac. Lavrentiev Ave. 3, 630090 Novosibirsk, Russia. · Meshalkin National Medical Research Center, Ministry of Public Health of the Russian Federation, Rechkunovskaya Str. 15, 630055 Novosibirsk, Russia.
Key points
From the paper's conclusion- Based on overall biological performance, Ag/Au heterostructures offering sustained antibacterial activity against planktonic bacteria, low cytotoxicity, and a cytokine profile that may be favorable for long-term integration are promising candidates for mesh implants.
Abstract
Ti-based mesh implants ("titanium silk" TiS and NiTi alloy) offer excellent tissue integration for oncologic surgery but lack intrinsic antibacterial activity. To address this, we fabricated Ag nanoparticles and Ag/M (M = Pt, Au) heterostructures on mesh surfaces via vapor-phase deposition, with noble metal sublayers functionalized by Ag nanofilms and nanoclusters. The modified implants were comprehensively characterized using SEM, AFM, XPS, and ICP-AES and evaluated through in vitro biocompatibility and antibacterial assays. The functional performance of the heterostructures is governed by two key phenomena: a chemical sublayer effect and a geometry-dependent substrate effect, which together dictate ion release kinetics. The Ag/Au interface sustains Ag+ release, whereas Ag/Pt drives rapid Ag+ dissolution. All heterostructures demonstrated effective antibacterial activity against planktonic S. aureus within 24 h. Key factors responsible for reduction in the more resilient P. aeruginosa on the Ag/M heterostructures were hypothesized. Cytotoxicity was dictated by the released ion cocktail: Ag/Pt/NiTi exhibited elevated toxicity which may be associated with the combined release of Ag+, Ni2+ and Pt2+, whereas all Ag/Au-based heterostructures showed excellent biocompatibility (viability 75-108%). Notably, gold sublayers were associated with elevated IL-10 secretion by 72 h. Based on overall biological performance, Ag/Au heterostructures offering sustained antibacterial activity against planktonic bacteria, low cytotoxicity, and a cytokine profile that may be favorable for long-term integration are promising candidates for mesh implants. Also, h-Ag/Pt/TiS showed an optimal balance between rapid antibacterial action and biocompatibility and therefore appears to be a suitable candidate for further in vivo investigation.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Bacterial Adhesion on Orthopedic Titanium Implants Coated with Silver-Doped Hydroxyapatite and Chitosan.Molecules (Basel, Switzerland) · 2026
- Handheld bioprinter-enabled in situ application of lignin-mediated green-synthesized AgNP-doped antibacterial GelMA-silk fibroin tissue adhesives: In vitro and ex vivo characterizations.International journal of biological macromolecules · 2026
- Silica nanoparticles-supported heterometallic Mo6Ag6 nanoclusters with enhanced antibacterial and antibiofilm performanceColloids and Surfaces B Biointerfaces · 2026
- Interfacial Chemistry‐Tailored Silica&Metal‐Based Heterostructures: from Rational Design to Antibacterial ApplicationsSmall · 2026
- Designing a heterostructure of graphite-like layer/TaOx on polyetheretherketone for selective antibacteria and osteogenesis.Regenerative biomaterials · 2026
- Bioinspired Mechano-Chemical Titanium Interfaces With Potential for Infection-Resistant Osseointegration.Small (Weinheim an der Bergstrasse, Germany) · 2026