Journal of clinical medicine

Extended Reality-Assisted Exoscopic Cervical Laminoplasty Integrating Intraoperative CT Navigation: A Technical Note on a Surgeon-Centered Virtual Workspace.

Kentaro Yamane, Wataru Narita, Shinichiro Takao, Hisakazu Shitozawa, Kazuhiro Takeuchi, Shinnosuke Nakahara

Department of Orthopaedic Surgery, National Hospital Organization Okayama Medical Center, 1711-1 Tamasu, Kita-ku, Okayama 701-1192, Okayama, Japan. · Department of Orthopaedic Surgery, Kameoka Municipal Hospital, 1-1 Shinonoda, Shino-cho, Kameoka 621-8585, Kyoto, Japan.

Key points

From the paper's conclusion
  • The system was used during exoscopic minimally invasive open-door laminoplasty through a 3-cm incision for multilevel cervical ossification of the posterior longitudinal ligament.

Abstract

Background/Objectives: In exoscopic spine surgery, the exoscopic image, navigation, and reference imaging occupy separate monitors. We describe a surgeon-centered extended reality (XR) workspace that places these sources in a single consumer head-mounted display, and report its measured video delay and its use in one clinical case. Methods: The stereoscopic output of a 4K three-dimensional (3D) exoscope and the intraoperative computed tomography (CT) navigation display were streamed over a dedicated local network to an Apple Vision Pro, together with four locally rendered reference layers; the exoscopic image was delivered at 1920 × 1080 in side-by-side stereoscopic format, that is, 960 × 1080 per eye. The delay of each streamed image relative to its native display was measured at the mirrored output with a 240-frames/s high-speed camera. The system was used during exoscopic minimally invasive open-door laminoplasty through a 3-cm incision for multilevel cervical ossification of the posterior longitudinal ligament. Results: The head-mounted exoscopic image lagged the native monitor by 30.8 ± 2.3 ms (range, 29.2-33.3 ms; n = 5), and the navigation image by 49.2 ± 1.9 ms (range, 45.8-50.0 ms; n = 5). The headset was worn for approximately 100 min without adjustment, malfunction, or reversion to the fallback monitor, and the procedure was completed without complication. Conclusions: The configuration was technically feasible with a small, measured delay. No ergonomic, workload, or clinical benefit is claimed.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers