Device-Derived Digital Biomarkers During SGLT2 Inhibitor Therapy: A Secondary Analysis of the ERASe Trial.
Ursula Rohrer, Katharina Ingrid Gölly, Martin Benedikt, Viktoria Santner, Martin Manninger, Simon Leiter, Johannes Widkal, Markus Stühlinger,
Department of Internal Medicine, Division of Cardiology, Medical University of Graz, Auenbruggerplatz 15, 8036 Graz, Austria. · University Clinic of Internal Medicine III/Cardiology and Angiology, Medical University of Innsbruck, Anichstraße 35, 6020 Innsbruck, Austria.
Key points
From the paper's conclusion- Conclusions: Ertugliflozin did not significantly alter individual device-derived parameters.
Abstract
Aims: Implantable cardioverter-defibrillators (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) provide continuous physiological data reflecting heart failure (HF) status and autonomic tone. Sodium-glucose cotransporter-2 inhibitors (SGLT2i) are core HF therapy, but it remains unclear whether their physiological effects are detectable using device-based remote monitoring. Following the ERASe trial, which showed reduced ventricular arrhythmias with ertugliflozin, this post-hoc analysis evaluated its impact on ICD/CRT-D-derived parameters. Methods: In this prospective, randomized ERASe post-hoc analysis, ertugliflozin was compared with placebo over 12 months in patients with ICD or CRT-D devices. Device-derived parameters including 24 h heart rate, night heart rate, heart rate variability, physical activity, and thoracic impedance were assessed at baseline and follow-up. Analyses were predefined as exploratory. Results: Forty-six patients (22 ertugliflozin, 24 placebo) had ≥1 analysable parameter. Paired longitudinal data were available for 29 patients for 24 h heart rate (HR), 19 for night HR, 17 for HR variability, and 9 for thoracic impedance. No significant between-group differences were observed in 24 h HR, physical activity or thoracic impedance. No significant differential longitudinal treatment effects were observed for the evaluated device-derived parameters, although estimates were imprecise because of limited parameter-specific data availability. Assessment of associations with arrhythmic and heart-failure outcomes was precluded by small event numbers and incomplete parameter-specific data. Conclusions: Ertugliflozin did not significantly alter individual device-derived parameters. Larger studies are needed to determine whether device-based physiological signals can identify treatment-related changes.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Hemodynamic changes in response to SGLT2 inhibitor treatment in ICD and CRT patients: insights from HeartLogic sensor dataEuropean Heart Journal · 2026
- Impact of SGLT-2 inhibitors on ventricular arrhythmias and cardiovascular mortality in patients with heart failure and ICD or CRT devices: a systematic review and meta-analysisEuropean Heart Journal · 2026
- Effect of dapagliflozin on electrocardiographic arrhythmic markers in patients with type 2 diabetes mellitus.British journal of clinical pharmacology · 2026
- Effects of luseogliflozin on sympathetic nerve activity:a dual assessment using iodine-123-metaiodobenzylguanidine and muscle sympathetic nerve activityEuropean Heart Journal · 2026
- Clinical outcomes and safety of SGLT2 inhibitor therapy in patients with a systemic right ventricle and heart failure: a pre-post analysis from the ACHIEVE-SGLT2i registryEuropean Heart Journal · 2026
- Complex antiarrhythmic effects of empagliflozin in non-ischemic HFrEF cardiomyocytes.Frontiers in pharmacology · 2026