From the paper's conclusion- Using the current standardized modification protocol, a PMEG can be constructed using the Zenith Alpha Thoracic Endovascular Graft while preserving the active fixation barbs for the treatment of cAAAs.
Abstract
Physician-modified endografts (PMEGs) represent an important treatment option for urgent and semi-urgent complex abdominal aortic aneurysms (cAAAs). The Zenith Alpha Thoracic Endovascular Graft (Cook Medical, Bloomington, IN, USA) is a common choice for PMEGs due to its low strut interference, albeit at the cost of requiring removal of the proximal fixation barbs for resheathing-a process that remains technically challenging. The current article details a step-by-step approach on how to modify an Alpha thoracic endograft without requiring removal of the proximal barbs. The Zenith Alpha Thoracic Endovascular Graft is fully unsheathed on a sterile back-table and completely detached from the delivery system by removing the blue rotational handle. The grey positioner is removed from the introducer sheath. Next, a 0.018-inch guidewire is introduced distally through the inner positioner and retrieved via the exposed grey handle to function as a trigger wire. Following the creation and reinforcement of the fenestrations, circular diameter-reducing ties are constructed and secured using an insertion tool. The endograft is placed back on the delivery system and the guidewire is passed through the endograft fabric at the distal end and through the insertion tool, which is then removed. At the proximal end, the trigger wire is again passed through the endograft fabric, and both the proximal bare alignment stent and the distal end of the endograft are secured to the delivery system using 2-0 Prolene sutures (Ethicon Inc., Somerville, NJ, USA). The endograft is subsequently resheathed through the distal end of the introducer sheath using a tourniquet-assisted resheathing technique and a cut-off tip from an introducer sheath to guide the endograft's passage through the valve. Using the current standardized modification protocol, a PMEG can be constructed using the Zenith Alpha Thoracic Endovascular Graft while preserving the active fixation barbs for the treatment of cAAAs.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Physician modified endografts in the aortic arch: History, techniques, outcomes, and considerations.Seminars in vascular surgery · 2026
- Physician-modified endografts using an endoskeleton ePTFE platform for complex aortic aneurysms requiring a single fenestration: a proof-of-concept study.CVIR endovascular · 2026
- Staged total arch repair for acute aortic dissection using physician-modified fenestrated thoracic endovascular aortic repair: a case report.AME case reports · 2026
- Endovascular Management With a Fenestration Technique of an Iatrogenic Thoracic Chronic Type B Aortic Dissection.Cureus · 2026
- First Report of Antegrade Transeptal Valve-in-Valve TAVR: Revisiting an Old Strategy.JACC. Case reports · 2026
- Clinically silent early proximal seal failure with suspected interval distal endograft displacement after hybrid zone 2 thoracic endovascular aortic repair in severe arch tortuosity: a case report with 1-year follow-up.Frontiers in medicine · 2026