Eng

Reflective Fiber-Tip Silica Microsphere Interferometer for Temperature Monitoring in Ethylene Glycol

Susan Ximena Mendoza-Trejo, Yamil Alejandro Zúñiga-Ávalos, Yadira Aracely Fuentes-Rubio, René F. Domínguez-Cruz, Arturo Alberto Castillo-Guzmán, Romeo de Jesús Selvas-Aguilar

Key points

From the paper's conclusion
  • We experimentally investigate a reflective fiber-tip interferometric sensor comprising a fused-silica microsphere at the end of a single-mode fiber and an external mirror, both immersed in ethylene glycol.

Abstract

We experimentally investigate a reflective fiber-tip interferometric sensor comprising a fused-silica microsphere at the end of a single-mode fiber and an external mirror, both immersed in ethylene glycol. The sensor was interrogated with a broadband source, an optical circulator, and an optical spectrum analyzer. For the five retained heating spectra, the selected peak shifted from 1546.35 nm at 25.2 °C to 1548.53 nm at 65.2 °C. The fitted wavelength slope was 51.6 pm/°C (R2 = 0.9862). Ten consecutive spectral acquisitions at 40.5 °C showed a standard deviation of 0.163 nm in the tracked peak position. The positive shift is inconsistent with a fixed-gap response governed solely by the negative thermo-optic coefficient of ethylene glycol. The individual contributions to the shift were not isolated. These measurements establish the feasibility of the proposed single-ended architecture for temperature monitoring in ethylene glycol, while highlighting the need for improved thermal referencing, mechanical stabilization, and repeated heating–cooling cycles before quantitative metrological claims can be made.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Reflective Fiber-Tip Silica Microsphere Interferometer for Temperature Monitoring in Ethylene Glycol — Science Skim