Diagnostic Performance of Angiography-Derived MPFFR for Predicting Hemodynamically Significant Coronary Stenosis: A Multicenter Validation Study.

Houng-Beom Ahn, Ju-Seung Kwun, Ki-Hyun Jeon, Wonjae Lee, Chang-Hwan Yoon, Hyun-Wook Chu, Won-Woo Seo, Donghoon Han,

Department of Internal Medicine, Seoul National University College of Medicine and Cardiovascular Center, Seoul National University Bundang Hospital, Seongnam 13620, Republic of Korea. · Department of Internal Medicine, Kangdong Sacred Heart Hospital, Hallym University College of Medicine, Seoul 05355, Republic of Korea.

Observational · n=189

Key points

From the paper's conclusion
  • This study evaluated the diagnostic performance of artificial intelligence (AI)-driven angiography-based FFR for detecting hemodynamically significant coronary stenosis, as compared with percentage diameter stenosis (%DS) from quantitative coronary angiography (QCA), using invasive FFR as the reference standard.

Abstract

Background and Objectives: Angiography-derived fractional flow reserve (FFR) methods offer noninvasive alternatives to wire-based FFR for functional assessment of coronary stenosis, but their diagnostic performance requires validation. This study evaluated the diagnostic performance of artificial intelligence (AI)-driven angiography-based FFR for detecting hemodynamically significant coronary stenosis, as compared with percentage diameter stenosis (%DS) from quantitative coronary angiography (QCA), using invasive FFR as the reference standard. Methods: In this prospective, multicenter study, 189 patients (211 lesions) undergoing clinically indicated angiography were enrolled. AI-driven angiography-based FFR and %DS were derived from angiographic images and compared with invasive FFR (cutoff ≤ 0.80). Diagnostic performance was assessed using AUROC, sensitivity, specificity, and predictive values, along with correlation analysis and analysis time. Results: AI-driven angiography-based FFR showed significantly greater diagnostic performance than %DS (per-patient AUROC: 0.87 vs. 0.63, p < 0.001; per-lesion AUROC: 0.89 vs. 0.65, p < 0.001). AI-driven angiography-based FFR demonstrated higher specificity (92.7%) and positive predictive value (86.7%) than %DS (64.2% and 51.9%, respectively). AI-driven angiography-based FFR correlated strongly with invasive FFR (r = 0.746), whereas %DS showed a weak inverse correlation (r = -0.250). Median analysis time for AI-driven angiography-based FFR was 20.9 s. A gray-zone hybrid strategy (thresholds 0.73-0.83) yielded 89.8% accuracy and could reduce invasive FFR use by 57.1%. Conclusions: AI-driven angiography-based FFR demonstrated high diagnostic accuracy, strong correlation with invasive FFR, and rapid analysis time. These findings support AI-driven angiography-based FFR as a pressure-wire-free tool for physiological assessment during coronary angiography.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers