Proactive Prediction of Alarms in Hemodialysis Machines: A Comparative Evaluation of AI Models for Clinical Prognostic Support.
Alessia Nicosia, Nunzio Cancilla, Marco Siino, Michele Passerini, Francesca Sau, Ilenia Tinnirello, Andrea Cipollina
Dipartimento di Ingegneria, Università degli Studi di Palermo, Viale delle Scienze Ed. 6, 90128 Palermo, PA, Italy. · Dipartimento di Ingegneria Elettrica, Elettronica e Informatica, Università di Catania, Via Santa Sofia 64, 95123 Catania, CA, Italy.
Key points
From the paper's conclusion- Results: Among the models tested, Random Forest emerged as the most robust predictive tool, achieving an F1 score higher than 0.85 for predicting alarms up to one minute in advance, with precision (~0.83), recall (~0.87), specificity (~0.99), AUROC (~0.99), and AUPRC (~0.94) values indicating strong performance, and stable performance across validation and test sets.
Abstract
Background/Objectives: Hemodialysis is a treatment for patients with kidney disease, yet it remains risk-prone due to potential technical or physiological instabilities. Dialysis machines rely on reactive safety mechanisms that trigger alarms only when predefined thresholds are reached. This often alerts clinical staff only after a complication has already occurred, contributing to potential delays in interventions. This study aims to develop a proactive prognostic framework capable of anticipating system instabilities before they transform into clinical adverse events. Methods: This work proposes a predictive approach for the proactive detection of patient- or process-related alarm conditions using ten machine learning models and one deep learning architecture. A dataset consisting of over two million records sampled at 0.5 s intervals from 95 dialysis sessions was assembled by merging machine logs of sensor readings and the binary machine status (alarm/normal). Results: Among the models tested, Random Forest emerged as the most robust predictive tool, achieving an F1 score higher than 0.85 for predicting alarms up to one minute in advance, with precision (~0.83), recall (~0.87), specificity (~0.99), AUROC (~0.99), and AUPRC (~0.94) values indicating strong performance, and stable performance across validation and test sets. A progressive decrease in performance was observed for longer prediction time windows. Conclusions: These findings demonstrate the feasibility of predicting dialysis-machine-generated alarms in advance using machine learning and deep learning models. Providing advance warning could give clinicians the opportunity to anticipate impending alarm conditions and, where appropriate, implement timely corrective actions, potentially improving treatment management and the operational reliability of renal replacement therapies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Artificial intelligence and machine learning applications in dialysis: Current applications, challenges, and future directionsClinica Chimica Acta · 2026
- Identification of acute kidney injury after non-cardiac surgery by machine learning leveraging a large electronic health record database.Scientific reports · 2026
- Early prediction of post-dialysis systolic and diastolic blood pressure using intradialytic data in Senegalese hemodialysis patients: a single-center retrospective cohort study.Frontiers in digital health · 2026
- Toward Transportable Acute Kidney Injury Prediction: An Explainable XGBoost Model with Temporal Validation Using MIMIC-IVmedRxiv · 2026
- Toward Explainable AI for Acute Kidney Injury Prediction in the ICU: Transparent Development, Internal Validation, and Robustness Auditing of Machine Learning ModelsInformation · 2026
- Development and validation of predictive models for the hemodialysis adequacy index (Kt/V) using multiple linear regression and machine learning algorithms.Frontiers in digital health · 2026