Comparison of Deep-Learning-Based Reconstruction of Knee MRI with Conventional Sequences in Patients with Knee Pain.
Rola Husain, Stephen Belmustakov, Etan Dayan, Guillermo Carbonell, Ajit Shankaranarayanan, Mingqian Huang, Idoia Corcuera-Solano
Diagnostic, Molecular and Interventional Radiology, Mount Sinai Health System, New York, NY 10029, USA. · Subtle Medical Inc., 883 Santa Cruz Ave, Menlo Park, CA 94025, USA.
Key points
From the paper's conclusion- To compare image quality and diagnostic performance among conventional sequences and an automated accelerated deep learning-based reconstruction (DLR) knee algorithm method in patients with knee pain.
Abstract
Background: Knee MRI is one of the most performed MR exams in outpatient settings and has a crucial role in detecting pathology. To increase access to advanced imaging and provide valuable diagnostic information to the clinician, reducing imaging acquisition time while preserving good image qualities is crucial. Deep learning-based reconstruction (DLR) algorithm (SubtleMR™) may be an alternative to conventional MRI with reduced scanning time while preserving image quality and diagnostic value. To compare image quality and diagnostic performance among conventional sequences and an automated accelerated deep learning-based reconstruction (DLR) knee algorithm method in patients with knee pain. Material and Methods: A total of 50 consecutive patients (26 men, mean age of 43.5 ± 10 years) with knee pain who underwent conventional (acquisition time: 13 min and 27 s) and accelerated DLR (acquisition time: 6 min and 45 s) knee MRI between November 2020 and December 2020 on a 1.5 T MRI system were prospectively enrolled. Two independent musculoskeletal radiologists compared the image quality artifacts and pathology using a 3-point Likert scale (1-3). Data were compared using paired Wilcoxon signed rank tests. Results: Forty-three patients with 86 knee MRI examinations were included. A comparison of conventional sequences and accelerated DLR by both readers revealed that there were no significant differences in overall diagnosis, image quality, and artifacts. Accelerated DLR knee MRI is comparable with conventional acquisitions, with an average time difference of 6 min and 42 s per patient (49.81%). Conclusions: Accelerated DLR knee MRI substantially reduces acquisition time (~49.81%) while maintaining equivalent diagnosis and image quality.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Evaluation of a Deep Learning Reconstruction Technique for Accelerated Brain MRI Scans.Tomography (Ann Arbor, Mich.) · 2026
- Rapid Musculoskeletal MRI in 2026: Clinical Integration of Deep Learning ReconstructionAmerican Journal of Roentgenology · 2026
- Diagnostic efficacy of deep learning-based denoising of low-field 0.55 T compared to conventional 3 T knee MRI.European journal of radiology · 2026
- Accelerated Renal T2-Weighted Turbo Spin-Echo Imaging with Deep Learning Reconstruction: Reduced Acquisition Time and Improved Image Quality in Patients with Renal TumorsBioengineering · 2026
- Clinical evaluation of deep learning-based acquisition and reconstruction techniques for accelerated 2D shoulder magnetic resonance imaging acceleration at 3 T.Skeletal radiology · 2026
- Deep-learning-assisted diagnosis for knee magnetic resonance imaging: Development and retrospective validation of MRNetPLoS Medicine · 2018