Clinical Repeatability and Inter-Method Agreement of VITA Easyshade Advance 4.0 and Rayplicker Cobra for Tooth Shade Determination: An Exploratory Cross-Sectional Study.
L Portero-Ruz, M Valor-Priego, A Martín-Vacas, M M Paz-Cortés, J Mena-Álvarez, C Rico-Romano
Faculty of Dentistry, Alfonso X el Sabio University, 28691 Madrid, Spain. · Faculty of Dentistry, Complutense University of Madrid, 28040 Madrid, Spain.
Key points
From the paper's conclusion- Easyshade showed moderate-to-substantial agreement with the visual comparator in selected regions, with the highest Kappa in the middle third of tooth 2.1 (0.619), whereas Rayplicker Cobra did not exceed fair agreement (maximum 0.362).
Abstract
Background: Instrumental shade determination may reduce some sources of subjectivity associated with visual shade selection, but device repeatability and interchangeability remain clinically relevant concerns. Objective: The aim of this study was to evaluate the reproducibility and clinical utility of two spectrophotometric devices (Easyshade and Rayplicker Cobra) for tooth shade measurement and selection. Methods: In total, 35 adults were evaluated at seven predefined maxillary tooth regions: the cervical, middle, and incisal thirds of teeth 2.1 and 2.3, and the middle third of tooth 2.6. Each instrumental measurement was repeated three times, with complete removal and repositioning of the device between readings. Visual shade selection using the VITA Classical guide was performed once per site by the same trained operator, whose normal colour vision had been confirmed before data collection. Teeth were kept hydrated, patients remained in the same position, and both devices were calibrated according to the manufacturers' instructions. Intra-device strict repeatability and inter-method agreement were assessed descriptively and using Cohen's Kappa, as appropriate. Results: Descriptive pooled strict repeatability was 52.2% for Easyshade and 53.1% for Rayplicker Cobra. Repeatability varied by tooth and region. Inter-device agreement was predominantly slight to fair (0.053-0.386). Easyshade showed moderate-to-substantial agreement with the visual comparator in selected regions, with the highest Kappa in the middle third of tooth 2.1 (0.619), whereas Rayplicker Cobra did not exceed fair agreement (maximum 0.362). Conclusions: Both spectrophotometers showed similar descriptive strict repeatability percentages; however, their low inter-device agreement indicates that their categorical shade outputs should not be considered interchangeable. The greater agreement of Easyshade with VITA Classical in selected regions reflects inter-method comparability rather than superior trueness, as no independent instrumental reference standard was used.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Agreement and Consistency of Visual and Digital Tooth Shade Selection: An In Vivo Comparison of Expert Practitioners and Senior Dental Students.Healthcare (Basel, Switzerland) · 2026
- Evaluation of Tooth Shade Selection With Conventional Methods and Artificial Intelligence-Integrated Methods: A Comparative Study.Cureus · 2026
- Impact of Material Aging and Time‐Dependent Factors on the Color Stability of Dental Shade Guides: A Spectrophotometric AssessmentColor Research & Application · 2026
- Digital Tooth Colour Acquisition: Agreement Between Spectrophotometric and Intraoral Scanner Measurements in Dental Students.Dentistry journal · 2026
- Influence of illumination conditions on color differences among visual, intraoral scanner, and spectrophotometric tooth shade determination methods: a prospective clinical study.BMC research notes · 2026
- Clinical Shade Matching of Dental Ceramics Using Digital Devices: A Systematic ReviewCurrent Oral Health Reports · 2026