From Ultrasound to Histopathology: Agreement Analysis and Depth Correction Model for Basal Cell Carcinoma Using a Portable 20 MHz High-Frequency Ultrasound Device
C R Popescu, Carmen Andrada Iliescu, Andreea Mihaela Truica, Liliana Gabriela Popa, Andrei Ludovic Poroșnicu, Irina Tudose, Carmen Boeru, M D Popescu
Carol Davila University of Medicine and Pharmacy · Clinical Emergency Hospital Bucharest
Key points
From the paper's conclusion- Conclusions: Twenty MHz HFUS demonstrated good agreement with histopathology for tumor depth assessment in BCC, with clinically acceptable variability.
Abstract
Background: High-frequency ultrasound (HFUS) represents a valuable non-invasive imaging modality for assessing cutaneous tumors. In basal cell carcinoma (BCC), preoperative estimation of tumor depth may support therapeutic decision-making and surgical planning. However, agreement between HFUS-derived and histopathologic depth measurements remains incompletely characterized, particularly with 20 MHz probes in routine clinical practice. Objectives: To evaluate agreement between 20 MHz HFUS and histopathology for BCC tumor depth using Bland–Altman analysis and to derive a preliminary correction equation to estimate histologic depth from HFUS measurements. Methods: This prospective observational pilot study included 15 patients with 16 histologically confirmed BCC lesions. All lesions underwent preoperative 20 MHz HFUS followed by surgical excision, with HFUS-derived tumor depth compared with histopathologic depth. Agreement was assessed using Bland–Altman analysis, and linear regression was performed to derive a preliminary correction equation. Results: Sixteen BCC lesions were analyzed. The mean difference between HFUS and histopathologic tumor depth was −0.07 mm, with 95% limits of agreement from −1.58 to +1.45 mm. HFUS and histopathologic depth measurements were highly correlated (R2 = 0.99). A correction equation was derived: estimated histopathologic depth (mm) = −0.52 + 1.10 × HFUS depth (mm). Conclusions: Twenty MHz HFUS demonstrated good agreement with histopathology for tumor depth assessment in BCC, with clinically acceptable variability. The proposed correction equation may improve interpretation of HFUS measurements; however, further validation in larger cohorts is required.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Hidden Depth, Incomplete Ablation: Basal Cell Carcinoma Recurrence After 20 MHz High-Intensity Focused Ultrasound (HIFU) Treatment.Clinical, cosmetic and investigational dermatology · 2026
- Beyond Breslow Thickness: Reproducibility and Histopathologic Concordance of Ultra-High-Frequency Ultrasound in Primary Cutaneous Melanoma.Diagnostics (Basel, Switzerland) · 2026
- Ex Vivo High-Frequency Ultrasound of the Skin in the Morphologic Assessment of Basal Cell Carcinoma: A Comparative Study.Diagnostics (Basel, Switzerland) · 2026
- Utility of High-Frequency Ultrasound in Preoperative Evaluation of the Thickness of Cutaneous MelanomaDiagnostics · 2026
- High-Frequency Ultrasound Assessment of Basal Cell Carcinoma: Correlations Between Histopathological Subtype, Vascularity, and Age/Sex DistributionCancers · 2026
- High-Frequency Ultrasound in High-Risk Cutaneous Squamous Cell Carcinoma.Actas dermo-sifiliograficas · 2026