Structural Design and Corrosion Response for GLAD-Engineered Titanium Thin Films for Medical Devices
Matteo Bertapelle, Joel Borges, Julia Claudia Mirza-Rosca, Filipe Vaz
Key points
From the paper's conclusion- Accordingly, this study systematically investigates the influence of growth geometry on the morphological, structural, and corrosion properties of titanium thin films, providing insights into the relationship between microstructural architecture and their protective performance.
Abstract
Five titanium thin films of comparable thickness (~1.0 ± 0.1 µm) were deposited by glancing angle deposition at geometrical deposition angles, α, of 0°, 20°, 40°, 60° and 85°. The reference sample (α = 0°) was prepared under conventional normal-incidence deposition conditions, resulting in vertically aligned columnar growth. The remaining four samples were deposited using a zigzag GLAD architecture, obtained by periodically rotating the substrate holder by 180° during deposition. All zigzag films were prepared using the same architecture, consisting of two zigzag bilayers, with the geometrical deposition angle being the only variable. Although titanium is widely recognized for its excellent corrosion resistance and biocompatibility, the protective performance of titanium coatings is strongly influenced by their microstructural characteristics, particularly porosity, surface roughness, and columnar morphology. GLAD offers a versatile approach to tailoring these features through precise control of the deposition geometry. Accordingly, this study systematically investigates the influence of growth geometry on the morphological, structural, and corrosion properties of titanium thin films, providing insights into the relationship between microstructural architecture and their protective performance. A clear transition was observed from dense, compact microstructures at low deposition angles to highly porous, anisotropic columnar morphologies at high angles, with a marked increase in surface roughness. All films crystallized in the hexagonal close-packed structure of titanium, with a progressive preferential growth transition from the thermodynamically favored (002) orientation at low angles to the kinetically driven (101) orientations at high angles, together with a reduction in overall crystallinity. Electrochemical analyses (open circuit potential, potentiodynamic polarization and EIS) demonstrated that corrosion resistance depends strongly on deposition geometry. Films produced at low (0–20°) and high (85°) deposition angles exhibited superior corrosion resistance due to the formation of stable passive oxide layers, whereas intermediate-angle films (40–60°) showed higher corrosion rates. These findings highlight the potential of tailored GLAD architecture for improving the performance of biomedical titanium coatings.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- 3D nanoscale design of oxide morphology, anisotropy and chemical heterogeneity by glancing angle physical vapor deposition.Nanoscale · 2026
- Magnetron sputtered Ta films with different double layer structures: Growth mechanism and corrosion behaviorJournal of Materials Research and Technology · 2026
- Fabrication and characterisation of Ti and DLC coatings on metamaterial-architecture-inspired 3D-printed polymer substratesThe International Journal of Advanced Manufacturing Technology · 2026
- DLC Coatings for Manufacturing Applications: Effects of Coating Structure on Mechanical and Tribological PerformanceLubricants · 2026
- Substrate Effects on the Corrosion Behavior of Tungsten- and Titanium-Doped Diamond-like Carbon Films on 316L Stainless Steel and Ti6Al4VProcesses · 2026
- Effect of Thin Film Deposition Method on Growth, Structural, and Optical Properties of ZnO:Al FilmsChemistrySelect · 2026