Chemistry

Electrochemical Detection of Hg(II) Using EASA-Derived Glutathione-Modified Mesoporous Silica Films

Deomila A. Basnig-Mamaoag, Ronel M. Mutia, Margarito C. Caadan, Rey Capangpangan

Key points

From the paper's conclusion
  • Films containing 5–15 mol% (3-mercaptopropyl)trimethoxysilane (MPTMS) were evaluated using charged redox probes, and the 10 mol% composition produced the highest electrochemical response.

Abstract

Mercury contamination threatens aquatic ecosystems and human health because of its toxicity, persistence, and potential for bioaccumulation. This study investigated an electrochemically assisted self-assembly (EASA)-derived glutathione-modified thiol-functionalized mesoporous silica film on a fluorine-doped tin oxide (FTO) electrode as a proof-of-concept interface for Hg(II) detection. Films containing 5–15 mol% (3-mercaptopropyl)trimethoxysilane (MPTMS) were evaluated using charged redox probes, and the 10 mol% composition produced the highest electrochemical response. FTIR analysis supported silica-network formation and the retention of glutathione-associated functional groups. Under the selected conditions of pH 5.0, a deposition potential of −0.70 V, and a deposition time of 5 min, the MSF-GSH/FTO electrode exhibited a linear response from 5 to 60 µM Hg(II), with an estimated calibration-based detection limit of 4.27 µM. Resolved oxidation peaks were observed for Hg, Pb, and Cd in a mixed-metal solution. The interface combines direct EASA film formation, thiol incorporation, and post-deposition glutathione modification; however, further sensitivity improvement, replication, stability testing, and real-sample validation are required.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers