Feasibility of an Intermittently Powered Bioelectronic Sensing System for Smart Knee Implants: A Coupled Electromagnetic–Thermal–Circuit Simulation Study
Albert Zsolt Barabas, Magdalena Miana Radu, Dorin Romeo Bica, Radu Ilinca, Ildiko Peter, Dana Rus, Oana Roxana Oprea
Key points
From the paper's conclusion- The present simulation framework therefore provides a quantitative basis for subsequent prototype development and experimental validation, while identifying the electrical, thermal, and energy-storage conditions that should be verified in future studies.
Abstract
Smart knee implants can provide objective functional information during postoperative follow-up, but long-term autonomous monitoring requires an implanted energy source capable of supporting sensing, processing, and wireless communication over extended periods. This study investigates an alternative architecture based on intermittent, physician-supervised operation, in which energy is transferred wirelessly immediately before a predefined measurement session. A 150 kHz inductive wireless power-transfer system was modeled in COMSOL Multiphysics to quantify the electrical output of an implanted receiver and evaluate electromagnetic losses and associated thermal effects in the receiver assembly, metallic tibial implant, bone, and surrounding soft tissue. The COMSOL-derived receiver port was subsequently reduced to a Thévenin equivalent and incorporated into an LTspice model comprising rectification, DC filtering, supercapacitor charging, regulated power delivery, and representative low-power implant electronics. The simulations showed that the received energy was sufficient to charge an 8 F, 2.7 V electric double-layer capacitor and subsequently maintain the modeled 3.0 V electronic supply for approximately 19.8 min after removal of the external power source. For the Main+Aux configuration, the maximum simulated temperature rise remained below 1 °C during 600 s of continuous excitation with a fixed 220 Ω receiver load. This thermal exposure was evaluated separately from the EDLC charging transient. All reported results for wireless power transfer, temperature rise, and post-charging operating time are based solely on simulations. These predictions support further evaluation of the proposed intermittent operating concept through prototype testing, including verification of electrical and thermal performance and assessment of mechanical integration, biocompatibility, and MR safety. The present simulation framework therefore provides a quantitative basis for subsequent prototype development and experimental validation, while identifying the electrical, thermal, and energy-storage conditions that should be verified in future studies.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Implantable Inductive Power Transfer: A Review of System Design, Safety, and ReliabilityElectronics · 2026
- Loss Analysis and Heat Optimization for a 700-mW Inductive Power Transfer System for Implanted Brain-Computer InterfaceIEEE Transactions on Power Electronics · 2026
- Sensor-integrated hip and knee prostheses: advances, challenges, and future perspectivesFrontiers in Bioengineering and Biotechnology · 2026
- Wireless Power and Data Transfer over Inductive Links for Biomedical Applications: Recent Advances, Challenges, and Limitations.Micromachines · 2026
- A Framework for Short-Range Wireless Power and Data Transfer in Miniaturized High-Power, High-Bandwidth Implants.Micromachines · 2026
- HDR-DTCP: High-Dynamic-Range Differential Tissue-Coupled Wireless Powering for Tunable Biphasic Stimulation in Injectable ImplantsarXiv · 2026