Flexible Nano-Platinum Microelectrode Arrays for Sensitive and Spatially Resolved Nitric Oxide Monitoring.
Umisha Siwakoti, Oshima Shah, Manoj Pandit, Kyle Rugg, Mark A DeCoster, Steven Amos Jones, Elisa Castagnola
Department of Biomedical Engineering, Louisiana Tech University, Ruston, LA 71272, USA.
Key points
From the paper's conclusion- These results support flexible nano-Pt/mPD MEAs for NO monitoring in complex biological environments and motivate experimental validation of spatially resolved measurements.
Abstract
Nitric oxide (NO) regulates neuronal signaling and neurovascular function, but its rapid diffusion, short lifetime, and reactivity complicate real-time detection in tissue. We developed flexible platinum microelectrode arrays (MEAs) modified with nanostructured platinum (nano-Pt) and a permselective poly(m-phenylenediamine) (mPD) coating for sensitive, selective NO monitoring. Nano-Pt electrodeposition reduced impedance approximately eightfold at 10 Hz and increased charge-storage capacity approximately 7.4-fold without increasing electrode dimensions. Simultaneous amperometric recordings from multiple electrode channels demonstrated the multisite sensing proof of concept of the flexible MEA platform. Amperometric calibration showed linear responses from 50 to 250 nM (R2 > 0.999), with a sensitivity of 243.16 ± 6.12 pA/µM and an operational detection limit of 50 nM. The mPD coating attenuated responses to ascorbic acid, uric acid, and dopamine. Sensors retained NO sensitivity (180 ± 8.6 pA/µM) in endothelial cell culture medium and detected transients following exposure of primary brain microvascular endothelial cells to copper-containing high-aspect-ratio structures (CuHARS), corresponding to peak NO-equivalent concentrations of 320 ± 50 nM. In mouse hippocampus, local NMDA/glycine administration evoked responses corresponding to estimated NO concentrations of 154 ± 31 nM. Finite-element simulations predicted detectable NO concentration gradients across neighboring electrode sites. These results support flexible nano-Pt/mPD MEAs for NO monitoring in complex biological environments and motivate experimental validation of spatially resolved measurements.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development and Innovative Applications of Tools for Monitoring Neuronal Electrical ActivityTheoretical and Natural Science · 2026
- A Flexible Carbon Nanofiber-Modified Multi-Channel Electrode Enables High-Sensitivity Electrochemical Detection of Multiple Neuroactive Molecules.Sensors (Basel, Switzerland) · 2026
- Zwitterionic Polymer Coating Enabled Chronic Dopamine Sensing and Electrophysiology Recording in Free-Moving Mice.Advanced healthcare materials · 2026
- Nanoplasmonic Aptasensing Enables Real-Time Optical Monitoring of Neurotransmitters in Living Brainstem Organoids.ACS nano · 2026
- Microfabricated Neural Biosensors for Detection of Neurotransmitters, Biomarkers, and Small Molecules: Emerging Trends on Self‐Sustained Systems and Energy HarvestingAdvanced Sensor Research · 2026
- Scalable Fabrication of 3D-Protruded Neural Microelectrodes Using Flexible PCB Technology for High-Fidelity Neural InterfacesACS Applied Electronic Materials · 2025