An Intelligent Wearable EMG Sensing Framework for Athlete Neuromuscular Monitoring and Performance Progression Assessment.
Kudratjon Zohirov, Sardor Boykobilov, Gulmira Pardayeva, Nilufar Akhmedova, Dilobar Ilmurodova, Iroda Uralova, Zavqiddin Temirov, Rashid Nasimov
Software and Hardware Support of Computer Systems, Karshi State Technical University, Karshi 180100, Uzbekistan. · Information Technologies, University of Information Technologies and Management, Karshi 180100, Uzbekistan.
Key points
From the paper's conclusion- Therefore, the results should be interpreted as preliminary, athlete-specific goodness-of-fit findings rather than evidence of generalizable predictive performance.
Abstract
Electromyography (EMG)-based sensing is an important tool for assessing neuromuscular activity and monitoring athlete development; its reliability depends on electrode placement, signal quality, and accurate identification of muscle activation periods. This study proposes an intelligent EMG sensing framework integrating preliminary electrode placement assessment, muscle activity detection, feature extraction, and regression-based progression prediction. A placement assessment indicated that positioning the electrode adjacent to the innervation zone produced the highest RMS under the tested conditions. A two-stage activity detection method based on clustering and probabilistic modeling achieved an average error of 1.5% and a temporal deviation of 19 ms. Nine time-domain EMG features extracted from the detected activity segments were used to characterize athlete progression and estimate the time required to reach a reference neuromuscular profile. Among the methods, Linear Regression provided the best fit to the data, obtaining R2 = 0.987 and RMSE = 4.21 and suggesting a predominantly linear relationship between the EMG-derived features and training duration within the dataset. However, these results were obtained from only six longitudinal observation periods for a single representative athlete, with each period represented by a 90-dimensional EMG feature vector derived from the ten movement classes. Therefore, the results should be interpreted as preliminary, athlete-specific goodness-of-fit findings rather than evidence of generalizable predictive performance. Validation using larger longitudinal cohorts and independent datasets is required. The proposed framework is compatible with future IoT-enabled wearable and edge-computing architectures; however, hardware-level implementation was beyond the scope of this study.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- A Hierarchical Three-Stage Feature Selection Strategy for Efficient and Generalizable sEMG-Based Joint Torque Estimation in Athletic Training and Rehabilitation.Sensors (Basel, Switzerland) · 2026
- How Stable Are Temporal EMG Parameters in Rowing? A Seven-Day Test-Retest Reliability Study Using Wearable sEMG.Sensors (Basel, Switzerland) · 2026
- Bioimpedance meets biomechanics: Wearable electrical impedance myography encodes fascicle and activation dynamics.Science robotics · 2026
- Surface-EMG Amplitude Differences Before and After Scheduled University Team Training Following a Standardized Warm-Up in Male Collegiate Volleyball Players.Journal of clinical medicine · 2026
- Real-time wearable biomechanics framework for sports injury prevention and rehabilitation optimizationScientific Reports · 2026
- A Standardized Methodology for EMG-Based Fatigue Assessment during Electrical Muscle StimulationCurrent Directions in Biomedical Engineering · 2026