Cold Atmospheric Plasma Disrupts Microbial Wound Bioburden: In Vitro, Porcine MRSA Biofilm, and Clinical Fluorescence Evidence from Bench to Bedside.
Steven Jeffery, Ahmad Wakaf, Lennart Marlinghaus, Sören Gatermann, Thomas Meyer, Joel Gil, Ivan Jozic, Stephen C Davis
Pioneer Wound Healing and Lymphoedema Centres (Pioneer-WHLC), Eastbourne BN23 8AS, UK. · Division of Surgery and Interventional Science, University College London (UCL), London WC1E 6BT, UK.
Key points
From the paper's conclusion- These findings support further controlled studies to define clinical benefit in wound care.
Abstract
Background: Biofilm-associated microbial burden is a major barrier to wound healing and is increasingly difficult to control in the era of antimicrobial resistance. This translational study evaluated the antimicrobial performance and tissue compatibility of a commercially available large-area cold atmospheric plasma (CAP) system (CPTpatch®/CPTcube®; Coldplasmatech GmbH, Greifswald, Germany) across in vitro, porcine, and clinical patient settings. Methods: CAP was tested against seven wound-relevant bacterial species, including multidrug-resistant strains, and the fungus Candida albicans in vitro, in two porcine MRSA-infected deep dermal wound models, and in a clinical service-evaluation cohort of 20 chronic lower-limb wounds in 14 patients; 17 wounds had evaluable longitudinal MolecuLight® (Toronto, ON, Canada) fluorescence series. Results: In vitro, CAP induced rapid multi-log killing across the bacterial panel and showed a marked antifungal effect against C. albicans. In porcine MRSA wounds, CAP reduced bacterial burden versus sham by 1.65, 1.89, and 2.04 log10 CFU/g on Days 4, 8, and 11, equivalent to 97.8%, 98.7%, and 99.1% reductions. In a 72-h post-inoculation MRSA biofilm model, the strongest 5×/week regimen achieved 2.51 log10 (99.69%) and 3.18 log10 (99.93%) reductions versus baseline after 2- and 4-min CAP exposures, respectively. Clinically, twice-weekly CAP was associated with a significant decline in MolecuLight® fluorescence grades over 5 weeks (p < 0.001); effective surface disinfection at 5 weeks was observed in 16/17 evaluable wounds. Conclusions: From bench to bedside, the investigated CAP system delivered rapid broad-spectrum antimicrobial activity, reduced MRSA burden in biofilm-infected porcine wounds, and was associated with clinically measurable suppression of wound surface bioburden while remaining tissue-sparing. These findings support further controlled studies to define clinical benefit in wound care.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Cold Atmospheric Plasma Improves Infected Wound Healing and Alters Microbial Oxidative Stress Responses and Host NRF2-Associated ResponsesAntioxidants · 2026
- Antibacterial and Antifungal Activity of a Cold Atmospheric Nitrogen Plasma Device in Porcine Burn and Excisional Wound Models: A Consolidated Pilot EvaluationApplied Sciences · 2026
- Cold Atmospheric Plasma Improves Healing Quality and Reduces Fibrosis in MRSA-Infected WoundsPlasma · 2026
- Cold Atmospheric Plasma in Biomedicine and Pharmacology: Multidisciplinary Perspectives on Therapeutic ApplicationsACS Pharmacology & Translational Science · 2026
- Cold atmospheric plasma and plasma-antibiotic combinations against ESKAPE pathogens: a systematic review of preclinical evidence.Frontiers in microbiology · 2026
- Cold Atmospheric Plasma-Aerosol Treatment of Equine Antibiotic-Resistant Wound Infection: A Preliminary Case Series.Animals : an open access journal from MDPI · 2026