Development and Validation of an SPE–LC–MS Method for the Determination of Epirubicin, Olaparib and Ribociclib in Human Serum
Monica Denisa Elena Popescu, Costel-Valentin Manda, O. Croitoru, Daniela-Maria Calucică, Johny Neamțu, Andrei Biţă, Amelia Maria Găman, Simona-Daniela Neamțu
University of Medicine and Pharmacy of Craiova
Key points
From the paper's conclusion- Results: The method demonstrated good linearity (r > 0.997) over the tested concentration ranges, along with adequate selectivity, precision, accuracy, recovery, and stability, fulfilling the ICH M10 guideline validation criteria.
Abstract
Background/Objectives: Epirubicin, Olaparib, and Ribociclib are widely used anticancer agents whose serum concentrations exhibit significant inter-individual variability, supporting the need for reliable and robust analytical methods suitable for pharmacokinetic evaluation and therapeutic exposure assessment. Variations in metabolism, drug–drug interactions, organ function, and treatment regimens may substantially influence systemic exposure, highlighting the importance of accurate quantification in clinical practice. This study describes the development and validation of a solid-phase extraction–liquid chromatography–mass spectrometry (SPE–LC–MS) method for the simultaneous quantification of these drugs in human serum. Methods: Sample preparation was performed using Oasis PRiME HLB® cartridges to ensure efficient clean-up, optimal recovery, and reduced matrix effects. Chromatographic separation was achieved using gradient elution with 0.1% formic acid and acetonitrile on a reversed-phase column, followed by single-quadrupole mass spectrometric (QDa) detection in the selected ion recording mode. The total run time was 13 min, enabling high-throughput analysis. Results: The method demonstrated good linearity (r > 0.997) over the tested concentration ranges, along with adequate selectivity, precision, accuracy, recovery, and stability, fulfilling the ICH M10 guideline validation criteria. No significant carry-over or interference from endogenous compounds was observed. Conclusions: Application to patient samples confirmed reliable performance in real clinical matrices and consistent quantification across different concentration levels. The proposed approach provides a potentially more accessible alternative in laboratories already equipped with LC-MS systems compared to LC-MS/MS platforms and can be applied in pharmacokinetic studies, representing a proof-of-concept for exposure assessment in oncology.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- QbD-guided HPTLC method development and validation for quantitative estimation of anticancer drugsJournal of Applied Pharmaceutical Research · 2025
- Simultaneous quantification of venetoclax, selinexor, voriconazole and fluconazole in human plasma by HPLC-MS/MS and its application in hematological malignancies.Frontiers in pharmacology · 2026
- Chromatographic Method for Simultaneous Quantification of Gemcitabine and Olaparib Recovered from Isolated Pancreatic TissueChemistry · 2026
- Development and Validation of an HPLC-MS/MS Method for Determining BMS-1166, GB-1107, and Enzalutamide in Rat Plasma and Its Application to Pharmacokinetic Studies.Drug design, development and therapy · 2026
- Analytical Quality by Design‐Based Reversed‐Phase High‐Performance Liquid Chromatography Method Development for the Simultaneous Determination of Ten Novel Anticancer Drugs and Establishment of Method Operable Design RegionSEPARATION SCIENCE PLUS · 2026
- Development and Validation of a Multianalyte LC-MS/MS Method for Quantification of BRAF and MEK Inhibitors and Cannabidiol in Cell Lysates and Cell Culture Media.Pharmaceutics · 2026