A Novel Dual-Function Face-Shield Prototype Integrating an On-Demand Hand-Sanitiser Reservoir: Design and Preliminary Benchtop Evaluation.
Pantelie Nicolcescu, Manuela Arbune, Aurel Nechita, Mădălina Nicoleta Matei, Ciprian Adrian Dinu, Anamaria Ciubara, Gabriel Valeriu Popa, Ada Stefanescu,
Faculty of Medicine and Pharmacy, Research Centre in the Field of Medical and Pharmaceutical Sciences (ReForm-UDJ), "Dunărea de Jos" University of Galați, Domnească Street, 800008 Galati, Romania. · Department of Ophthalmology, University of Medicine and Pharmacy of Craiova, 200349 Craiova, Romania.
Key points
From the paper's conclusion- We designed and performed a preliminary benchtop evaluation of a proof-of-concept face-shield prototype integrating an on-demand disinfectant reservoir.
Abstract
BACKGROUND AND OBJECTIVES: Rapid hand hygiene is difficult to maintain during uninterrupted patient care. We designed and performed a preliminary benchtop evaluation of a proof-of-concept face-shield prototype integrating an on-demand disinfectant reservoir. MATERIALS AND METHODS: A dual-function wearable system was built from a PETG visor (230 × 230 mm, 0.20 mm) and modular reservoirs (25, 50, 75, 100 mL). Five complete prototype units were fabricated under the same manufacturing conditions: two separate physical units with 100 mL reservoirs (Units A and B) and one unit each with 75, 50, and 25 mL reservoirs. Functional and mechanical testing was performed across the two 100 mL units, with shared and unit-specific endpoints; chemical compatibility was evaluated using dedicated PETG specimens, while ergonomic assessment was limited to biomechanical modelling. Disinfectant density (Mikrozid AF Liquid) was measured (0.875 ± 0.002 g/mL, n = 5) to convert dispensed mass to volume. The valve delivered 3.08 ± 0.21 mL/activation (CV 6.8%, n = 60; 30/unit, single fill, no refilling). RESULTS: The system weighed 103.3 g empty and 190.8 g filled with 100.0 mL of disinfectant (87.5 g disinfectant mass). Recalculated cervical torque (τ = m × g × r; baseline: 0.084 N·m) was 0.118 N·m for the empty-reservoir configuration (+40.6%) and 0.252 N·m for the filled configuration (+199.9%). No continuous leakage occurred (mass loss < 0.1 g), retention exceeded 10 N in all directions, and PETG remained stable after 14-day immersion and 50 wiping cycles. Estimated cost was €10-20/unit. CONCLUSIONS: This proof-of-concept supports the technical feasibility of integrating a face-shield platform with immediate access to hand sanitisation in one reusable device, without establishing standards-based protective performance. The small number of separate physical prototypes used for functional testing (two 100 mL units), together with the absence of microbiological, optical-standard and human-factor validation, limits generalisability. Human-participant, microbiological, ergonomic and clinical usability studies are required before clinical implementation. A conceptual touchless valve is proposed for future work.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Detection Efficacy of a Novel Disposable Dental Protective Device versus Medical Goggles in Reducing Splatter Exposure during Tooth Preparation: A Randomised Controlled Trial.Oral health & preventive dentistry · 2026
- Applicator-Based Versus Bulk-Bottle Delivery of the Same Alcoholic Chlorhexidine: A Single-Center, Non-Randomized Pilot Micro-Costing Study of Two Surgical Skin-Antisepsis Workflows.Healthcare (Basel, Switzerland) · 2026
- Quantifying User-Applied Forces on Prefilled Syringe Luer Cone: A Human Factors Ergonomics Study.PDA journal of pharmaceutical science and technology · 2026
- From cabinet to PlasmaBag: validating mobile endoscope storage with plasma technology.Microbiology spectrum · 2026
- In-Situ Disinfection of Hospital Privacy Curtains Using a Next-Generation Aerosolized Hydrogen Peroxide SystemmedRxiv · 2026
- Real-Time Wristband Monitoring of Hand Hygiene Technique Using Motion Sensors and Machine Learning.Journal of visualized experiments : JoVE · 2026