From the paper's conclusion
  • Conclusions: The wearable system recorded task- and region-dependent inter-sensor tilt patterns during TLSO wear, with the largest exploratory estimates between Th12 and L4 during crouching and forward bending.

Abstract

Background: Thoracolumbar orthoses are used as one component of non-operative spinal injury management, but clinical evidence regarding their benefit is inconsistent and their segment- and task-specific biomechanical effects remain insufficiently quantified. Purpose: This study aimed to quantify region- and task-specific changes in sagittal inter-sensor tilt during static postures with and without one individually fitted thoracolumbosacral orthosis (TLSO). Study Design/Methods: This exploratory repeated-measures laboratory study included 17 healthy adults (10 male, 7 female; mean age, 31.2 ± 10.9 years). Five wireless triaxial accelerometers were positioned at Th10, Th12, L2, L4, and S1. Adjacent inter-sensor tilt differences and the Th10-S1 difference were measured during sitting, crouching, forward bending, and backward bending with and without the TLSO. Paired conditions were analyzed using Wilcoxon signed-rank tests. Global Benjamini-Hochberg false discovery rate correction was applied across 20 comparisons, and matched-pairs rank-biserial correlations (r_mrb) and bootstrap 95% confidence intervals were calculated. Results: After global false discovery rate correction, no comparison reached statistical significance (all p_FDR > 0.05). Exploratory estimates indicating lower signed inter-sensor tilt with the TLSO were observed at L2-L4 during crouching (median 14.5° without vs. 1.6° with TLSO; r_mrb = 0.765, 95% CI 0.40-0.99; p_FDR = 0.0697), L2-L4 during forward bending (20.6° vs. 5.1°; r_mrb = 0.712, 95% CI 0.28-1.00; p_FDR = 0.0697), Th12-L2 during forward bending (12.1° vs. 0.7°; r_mrb = 0.660; p_FDR = 0.0697), and Th12-L2 during crouching (6.1° vs. 2.0°; r_mrb = 0.595; p_FDR = 0.1017). Conclusions: The wearable system recorded task- and region-dependent inter-sensor tilt patterns during TLSO wear, with the largest exploratory estimates between Th12 and L4 during crouching and forward bending. Because no comparison remained statistically significant after correction, the findings are hypothesis-generating and require validation in larger patient cohorts and with additional orthosis models.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers