Antibiotics

Highly Sensitive Determination of Tiamulin in Meat Products by Lateral Flow Immunoassay with Magnetic Preconcentration

Anatoly V. Zherdev, Nadezhda A. Taranova, Elena A. Zvereva, Alisa A. Bulanaya, Angelina A. Aslamova, Boris B. Dzantiev

Key points

From the paper's conclusion
  • The feasibility of applying the approach to preconcentration of tiamulin from real food matrices (chicken, beef, and egg) is demonstrated.

Abstract

Background: Magnetic particles are a unique marker for immunochromatographic test systems, enabling the concentration of the analyte from large-volume samples. The proposed test systems differ in the markers used (gold nanoparticles and magnetic particles) and the analysis format (direct and indirect marker introduction into the analytical zone). Method: The most effective test system is based on the use of magnetic particles as a carrier of specific anti-bodies, followed by colorimetric detection of the analytical signal by introducing a colloidal gold conjugate with anti-species antibodies. Results: For the developed test system, a visual detection limit of 0.13 ng/mL and an instrumental detection limit of 89 pg/mL (27 pg/mL with concentration) were achieved. Enhancement with colloidal gold allowed us to reduce detection limits by 11 times, and preliminary concentration by up to 37 times. The feasibility of applying the approach to preconcentration of tiamulin from real food matrices (chicken, beef, and egg) is demonstrated. The analyte detection rate was 94–108%.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers