Frontiers in surgery

A "stepped" flexible vacuum-assisted ureteral access sheath improves stone clearance efficiency: experimental results from an ex vivo porcine model.

Daibing Lv, Ming Jin, Xi Meng, Weizhong Zhang, Shanfang Yuan, Yang Ding, Weixi Wang, Jianghua Yang,

Department of Urology, Beijing University of Chinese Medicine Third Affiliated Hospital, Beijing, China.

Key points

From the paper's conclusion
  • We simulated f-URL stone extraction under continuous irrigation and vacuum suction, comparing the SFV-UAS (ureteral segment 12/14 Fr, cystourethral segment 16/18 Fr) with a conventional flexible vacuum-assisted ureteral access sheath (FV-UAS)(12/14 Fr).

Abstract

OBJECTIVE: To evaluate the impact of a novel "stepped" flexible vacuum-assisted ureteral access sheath (SFV-UAS) on stone clearance efficiency during flexible ureteroscopy lithotripsy (f-URL). METHODS: Human upper urinary tract stone fragments (< 4 mm) were placed into ex vivo porcine renal pelves. We simulated f-URL stone extraction under continuous irrigation and vacuum suction, comparing the SFV-UAS (ureteral segment 12/14 Fr, cystourethral segment 16/18 Fr) with a conventional flexible vacuum-assisted ureteral access sheath (FV-UAS)(12/14 Fr). With intrarenal pressure strictly controlled (< 30 mmHg) and constant vacuum suction (01 step), irrigation flow rates were evaluated at 40 mL/min, 50 mL/min, 60 mL/min, 70 mL/min, 80 mL/min, and 90 mL/min. Stone evacuation times were recorded. RESULTS: (1) Sheath type and irrigation flow rate exhibited significant main effects on clearance time [F(1,408) = 189.231, P < 0.01, effect size 31.7%; and F(5,408) = 38.721, P < 0.01, effect size 32.2%, respectively]. A significant two-way interaction was also observed [F(5,408) = 8.372, P < 0.01, effect size 9.3%]. (2) Simple main effects revealed: First, across all corresponding irrigation rates, the SFV-UAS group achieved significantly shorter clearance times than the FV-UAS group (mean differences: MD = -21.598, -18.099, -8.172, -11.129, -9.762, -4.883; all P < 0.05). Second, within the SFV-UAS group, rates ≥ 70 mL/min resulted in significantly shorter clearance times vs. the 40 mL/min baseline (mean differences:MD = -9.393, -9.771, -9.084; all P < 0.05). Although comparisons between 60 mL/min and higher rates were non-significant, 70 mL/min was the lowest evaluated irrigation rate that demonstrated a statistically significant reduction in clearance time compared to the 40 mL/min baseline. Third, within the FV-UAS group, clearance times at ≥ 60 mL/min were significantly shorter than at 40 and 50 mL/min (mean differences:MD = -19.664, -12.370; all P < 0.001), with no significant differences among the ≥ 60 mL/min groups, indicating its plateau at ≥ 60 mL/min. CONCLUSION: Within the parameters of this ex vivo model,the SFV-UAS significantly reduces the measured stone clearance time during f-URL compared to the FV-UAS under identical irrigation conditions. For the SFV-UAS group, targeting an irrigation flow rate of ≥ 70 mL/min consistently optimizes stone extraction efficiency relative to baseline conditions.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

A "stepped" flexible vacuum-assisted ureteral access sheath improves stone clearance efficiency: experimental results from an ex vivo porcine model — Science Skim