Safety and efficacy of a novel four-hook needle versus hook wire for CT-guided preoperative localization of pulmonary nodules: a propensity score-matched analysis.
Rui Zhang, Yue Hu, Fangfang Tian, Xiaohui Zhao, Pengtao Bao, Haipeng Yu
Department of Interventional Therapy, Tianjin Medical University Cancer Institute & Hospital, Tianjin, China. · Department of Neurosurgery, Aviation General Hospital, Beijing, China.
Key points
From the paper's conclusion- Background To compare the safety and effectiveness of a novel four-hook needle with conventional Hook wire for CT-guided preoperative localization of pulmonary nodules and to identify factors associated with localization-related complications.
Abstract
BACKGROUND: To compare the safety and effectiveness of a novel four-hook needle with conventional Hook wire for CT-guided preoperative localization of pulmonary nodules and to identify factors associated with localization-related complications. METHODS: A retrospective study was conducted in patients who underwent CT-guided localization followed by video-assisted thoracoscopic surgery between January 2024 and June 2025. Propensity score matching yielded 248 matched pairs. Post-matching comparisons were performed using paired statistical methods. Generalized estimating equation models with the same reduced set of clinically relevant covariates were used for pulmonary hemorrhage and pneumothorax, with matched pair specified as the clustering variable. RESULTS: Technical localization success was achieved in all patients. Tip-to-nodule distance was comparable between the Hook wire and four-hook needle groups (4.82 ± 4.91 mm vs. 4.74 ± 4.22 mm, P = 0.842). The four-hook needle shortened localization time (7.69 ± 3.18 min vs. 8.76 ± 3.22 min, P < 0.001). Pulmonary hemorrhage (8.1% vs. 13.7%, P = 0.054) and pneumothorax (10.1% vs. 16.1%, P = 0.067) occurred less frequently in the four-hook group, although the differences did not reach statistical significance. Marker dislodgement was significantly less frequent with the four-hook needle (0.4% vs. 3.2%, P = 0.039), and post-localization VAS pain scores were lower (2.56 ± 1.28 vs. 4.00 ± 1.21, P < 0.001). In matched-pair clustered multivariable analyses, greater distance from the pleura, greater puncture depth, and more needle adjustments were associated with pulmonary hemorrhage, while more needle adjustments and prone positioning were associated with pneumothorax. CONCLUSIONS: Both localization techniques provided accurate preoperative localization of pulmonary nodules. The four-hook needle shortened localization time, reduced marker dislodgement, and was associated with lower post-localization VAS pain scores. Pulmonary hemorrhage and pneumothorax were numerically less frequent with the four-hook needle, but these differences were not statistically significant after accounting for the matched-pair structure.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Clinical application of CT-guided lung nodule localization needles in preoperative localization of small pulmonary nodulesJournal of Cardiothoracic Surgery · 2026
- Preoperative computed tomography-guided localization of pulmonary nodules: comparison between tailed coil and suture anchor localization.Wideochirurgia i inne techniki maloinwazyjne = Videosurgery and other miniinvasive techniques · 2026
- Comparison of CT-guided suture hook anchor and indocyanine green localization for multiple pulmonary nodules.Frontiers in oncology · 2026
- Intraoperative titanium clip localization versus preoperative medical glue localization for pulmonary nodules: a prospective cohort study.World journal of surgical oncology · 2026
- CT-guided Preoperative Localization of Impalpable Pulmonary Nodules: Comparison Between Two Devices.In vivo (Athens, Greece) · 2026
- Nodule Dye-Localization Using Interventional Fluoroscopy for Image-Guided Sublobar Lung Resection.Thoracic cancer · 2026