Three decades of cochlear implant failure and multiple reimplantation: a retrospective cohort study.
Luluh Alsughayer, Nezar Hamed, Thamer Alzahrani, Abdullah Alsarar, Rahma Sweedy, Turky AlHarbi, Wafaa S Taishan, Afrah Alshalan,
Department of Otolaryngology-Head and Neck Surgery, College of Medicine, King Saud University, Riyadh, Saudi Arabia. · Department of Otolaryngology-Head and Neck Surgery, King Faisal Specialist Hospital and Research Centre, Riyadh, Saudi Arabia.
Key points
From the paper's conclusion- Conclusion CI failure had a prevalence of 2.41% over more than three decades.
Abstract
INTRODUCTION: Cochlear implantation (CI) is a highly reliable intervention for sensorineural hearing loss; however, device failure may occur, limiting patient benefit unless revision surgery is performed. This study evaluated the long-term institutional prevalence and causes of CI failure, and assessed complications, audiological and speech outcomes, and their associated factors following multiple revision surgeries. MATERIALS AND METHODS: A retrospective cohort study identified 91 failed CI cases among 3,777 CIs performed at a single tertiary referral center between January 1993 and November 2024. Audiological and speech outcomes, including Pure Tone Average (PTA), Speech Reception Threshold (SRT), Speech Discrimination Score (SDS), Categories of Auditory Performance (CAP), and Speech Intelligibility Rating (SIR), were analyzed using Mixed Models for Repeated Measures (MMRM) with restricted maximum likelihood (REML) estimation. RESULTS: The overall prevalence of CI failure was 2.41%, occurring predominantly in pediatric patients (96.7%). Hard device failure was the most common cause (68.8%), followed by medical (15.0%), soft (13.8%), and surgical (2.5%) failures. Revision surgery was performed once in 80 cases, twice in 8 patients, and three times in 2 cases. Reimplantation successfully preserved audiological performance across successive revisions, while CAP and SIR scores demonstrated significant additional improvement after the first revision compared with the initial implantation (p = 0.037 and p < 0.001, respectively). Reimplantation using the same electrode model was associated with better SDS after the first revision (p = 0.035), and better SRT (p = 0.022) and SIR (p = 0.001) after the second revision. An interval of >3 years between surgeries was associated with significantly higher post-revision SIR scores after both the first (p = 0.040) and second (p = 0.004) revisions. CONCLUSION: CI failure had a prevalence of 2.41% over more than three decades. Reimplantation is a salvage procedure that successfully restores auditory function after device failure, allowing for continuous language development. Electrode selection and revision strategy should remain tailored to each case.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Data-Driven Multiparameter Predictive Analysis of Cochlear Implant Rehabilitation Outcomes.Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2026
- Long-term multidimensional outcomes of China’s national cochlear implant assistance programme over twenty-one yearsScientific Reports · 2026
- Establishing reliable neural health measures in adult cochlear implant users.Frontiers in neuroscience · 2026
- A Multicenter Study of Cochlear Implantation in Adults With Prolonged Unilateral Deafness.The Laryngoscope · 2026
- Quantifying Soft‐Surgery in Cochlear Implantation: Multimodal Data From 30 International SpecialistsThe Laryngoscope · 2026
- Non-usage rates after paediatric cochlear implantation: A systematic review and meta-analysis.International journal of pediatric otorhinolaryngology · 2026