CT-based navigation for acetabular cup placement in supine and lateral positions: a simulated surgery study using patient-specific 3D-printed pelvic models.
Lanyu Liang, Hassan M Nemati
Ortoma AB, Gothenburg, Sweden.
Key points
From the paper's conclusion- The aim of this study was to evaluate the accuracy and precision of a CT-based, AI-driven navigation tool for acetabular cup placement in supine and lateral patient positions, using a patient-specific 3D-printed pelvic model with an embedded acetabular cup rim as a physical measurement reference.
Abstract
PURPOSE: Accurate acetabular cup placement is critical for the success of total hip arthroplasty (THA). The aim of this study was to evaluate the accuracy and precision of a CT-based, AI-driven navigation tool for acetabular cup placement in supine and lateral patient positions, using a patient-specific 3D-printed pelvic model with an embedded acetabular cup rim as a physical measurement reference. METHODS: Twenty-six patient-specific pelvic models were generated from preoperative CT data. A virtual acetabular cup was added to each pelvic mesh, cropped to the rim, and 3D-printed together with the bone, yielding a physical cup reference embedded in the printed pelvis. The ground-truth cup orientation and position were defined within the planning application by overlaying a cup component template on the embedded rim. Each model was tested once in supine and once in lateral position in a paired design. Ten cases were repeated three times per position to assess intra-session precision. Outcomes were surface matching root-mean-square error (RMSE), cup anteversion, inclination, and depth errors. Analyses included two one-sided tests (TOST) for equivalence (±0.5 mm), Bland-Altman agreement, Levene's test, and intraclass correlation coefficients (ICC). RESULTS: Mean alignment-point RMSE was 0.26 mm (supine) and 0.28 mm (lateral); mean independent test-point RMSE was 0.30 mm in both positions, with TOST confirming equivalence within ±0.5 mm. Mean absolute anteversion error was 0.67° (supine) and 0.44° (lateral), with comparable inclination errors. Mean absolute cup depth error remained well below ±3 mm. Bland-Altman analyses showed narrow 95% limits of agreement between positions. All ICC(3,1) values for cup orientation exceeded 0.99 in both positions. CONCLUSIONS: The evaluated navigation tool achieved sub-millimeter surface matching, small cup placement errors, and excellent intra-session precision in both supine and lateral positions. The embedded-rim phantom enabled direct measurement of cup orientation and seating depth on the printed model and is applicable to other CT-based navigation systems.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- TP4.5 Precision Without Navigation: Standardized C-Arm Positioning Improves Accuracy in Periacetabular Osteotomy Surgical ModelJournal of Hip Preservation Surgery · 2026
- Accuracy and risk factors for poor cup orientation in total hip arthroplasty: a comparison between supine and lateral registration using a portable navigation system.Archives of orthopaedic and trauma surgery · 2026
- TP4.3 Accuracy of Patient Specific Guide to Achieve Pre-Operatively Templated Correction of Hip Dysplasia with PAOJournal of Hip Preservation Surgery · 2026
- Accurate Acetabular Component Positioning in Total Hip Arthroplasty Using Preoperative CT and Anatomical Landmarks.Journal of clinical medicine · 2026
- Robotic-referenced automated measurement of acetabular cup orientation on postoperative CT: development and internal validation.Journal of robotic surgery · 2026
- Smartwatch-based pelvic tracking system for augmented reality navigation in total hip arthroplasty: a preliminary clinical study.Arthroplasty (London, England) · 2026