Frontiers in transplantation

Prolonged ex vivo lung perfusion with corrected perfusate: functional and structural benefits in swine.

Justin Issard, Julia Mercier, Jean-Baptiste Ménager, Guillaume Fadel, Yuya Nobori, Melwan Izem, Maria-Rosa Ghigna, Elie Fadel,

Department of Thoracic Surgery and Heart-Lung Transplantation, Marie Lannelongue Hospital, Université Paris-Saclay, Le Plessis Robinson, France. · Hypertension Pulmonaire: Physiopathologie et Innovation Thérapeutique (HPPIT), Hôpital Bicêtre (AP-HP), Hôpital Marie Lannelongue, FHU André Cournand, ERN-LUNG, Université Paris-Saclay, INSERM UMR_S 1358, Le Plessis Robinson, Le Kremlin-Bicêtre, France.

Randomised · n=6

Key points

From the paper's conclusion
  • Maintaining physiological perfusate composition improved lung function during 12-hour EVLP in swine; however, no post-transplant benefit was observed. Further studies are warranted to assess the impact of this strategy during prolonged EVLP and to compare it with alternative approaches.

Abstract

BACKGROUND: During ex vivo lung perfusion (EVLP), perfusate pH, electrolyte, and glucose concentrations often deviate from physiological ranges. We evaluated the effect of a corrective solution on maintaining these parameters within physiological limits during 12 h of EVLP in a porcine model. METHODS: Twelve porcine lung blocks were harvested and cold-preserved before undergoing 12 h of normothermic EVLP. Lung blocks were randomly assigned to two groups based on the perfusate replacement strategy. All lungs were perfused with Steen solution, mean priming volume was 1,525 ± 340 and 1,325 ± 232 mL, with 100 mL (5%-9%) replaced every two hours by either fresh Steen solution (standard group, n = 6) or a modified solution designed to restore physiological composition (corrected group, n = 6). Perfusate electrolytes, glucose, cytokines, pH, partial oxygen pressure (PO₂), pulmonary vascular resistance (PVR), and compliance were measured. After EVLP, left lungs were transplanted during 4 h into donor-related pigs, while right lungs were analyzed by histology and electron microscopy. RESULTS: Perfusate correction successfully maintained electrolyte, pH, and glucose levels within target ranges during EVLP. The corrected group demonstrated higher lung compliance and oxygenation, reduced edema, more hyaline membranes, and a greater density of Kohn's pores after EVLP. PVR was similar between groups, and no significant difference in ΔPO₂ was observed post-transplantation. CONCLUSION: Maintaining physiological perfusate composition improved lung function during 12-hour EVLP in swine; however, no post-transplant benefit was observed. Further studies are warranted to assess the impact of this strategy during prolonged EVLP and to compare it with alternative approaches.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers