Frontiers in PhysicsOpen accessPenning ionization

Numerical and experimental analysis of registration of small ammonia impurities using the PLES method

Jie Zhang, Almaz I. Saifutdinov, Zijia Chu, Chen Zhou, Chengxun Yuan, Zhongxiang Zhou

Harbin Institute of Technology

Key points

From the paper's conclusion
  • Discussion The results can provide significant reference for the detection of trace impurities in gases based on PLES method.

Abstract

Introduction This paper investigates the applicability of detecting ammonia impurities in plasma electron spectroscopy (PLES) through both numerical and experimental approaches. Methods A self-consistent plasma hybrid fluid-kinetic model is used to accurately describe the formation of characteristic peaks associated with fast electrons in inelastic collision reactions observed in experiments. The model is introduced to calculate the minimum of detectable impurity density at low pressure. Results Numerical simulations based on the 1D model predict an idealized theoretical detection limit of approximately 5 ppm for ammonia impurities; however, this value should be regarded as a model-derived theoretical lower bound rather than an experimentally confirmed figure. In practice, factors such as signal-to-noise ratio, probe contamination, and measurement noise make it challenging at this stage to achieve direct and stable detection at such low densities. In addition, the effect of ammonia densities on the electron energy distribution function is studied. At low densities, Penning ionization of metastable helium atoms correlates linearly with ammonia density. This process increases the energy transfer rate and significantly enhances characteristic peak intensities. Conversely, at high densities, metastable species undergo rapid depletion. Frequent inelastic collisions between electrons and ammonia molecules dissipate energy into vibrational modes, ultimately driving the electron energy distribution function into a saturated state dominated by collisional quenching. Discussion The results can provide significant reference for the detection of trace impurities in gases based on PLES method.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers

Numerical and experimental analysis of registration of small ammonia impurities using the PLES method — Science Skim