Cardio-respiratory-gated proton pencil beam scanning for cardiac ablation: workflow and end-to-end validation.
Nicolas Depauw, Felicitas Bijari, Ethan Cascio, Evan Driscoll, Adriano Garonna, Jagdip Kang, Helen Deng, Hyeongseok Kim,
Mass General Brigham, Harvard Medical School, Boston, MA, United States. · EBAMed SA, Geneva, Switzerland.
Key points
From the paper's conclusion- These findings are valuable for establishing a workflow for cardiorespiratory-gated STAR treatment with CardioKit.
Abstract
BACKGROUND: Stereotactic Arrhythmia Radiotherapy (STAR) offers a non-invasive alternative for ventricular tachycardia ablation, yet motion management remains a critical challenge. While respiratory gating is established, cardiac gating requires a dedicated workflow and gating device. METHODS: This study establishes and validates an end-to-end workflow for cardiorespiratory gated proton STAR in six healthy swine. Key innovations include a novel pencil-beam scanning (PBS) controller enabling mid-spot beam pausing and resumption, and the integration of CardioKit, an ultrasound-based image-guidance system for real-time soft-tissue tracking. The workflow was evaluated by comparing single-gated (respiratory only) and dual-gated (cardiorespiratory) delivery strategies, assessing system latency, dosimetric accuracy on dynamic phantoms, and motion tracking fidelity against 4D-CT benchmarks. RESULTS: The upgraded PBS system demonstrated a gate response latency of <550 µs, effectively mitigating interplay effects in dynamic phantom measurements through gated beam deliveries. CardioKit provided respiratory motion traces highly correlated with 4D-CT data (Pearson r > 0.97) and responded to arrhythmic events within 5 ms. In vivo, dual-gating restricted delivery to the diastolic window (70-100% R-R), which resulted in a reduction in the Internal Target Volume (ITV) expansion by a mean of 45% compared to single-gating (19% vs. 64%), thereby substantially sparing healthy myocardium. Although dual-gating increased total beam-on time, the mid-spot interrupt capability ensured robust dose accumulation without treatment aborts. CONCLUSION: The workflow for imaging, treatment planning, and cardiorespiratory-gated beam delivery was viable for treating all six subjects. The PBS system responded to gating commands < 550 µs and CardioKit responded to irregularities in heartbeat in <5 ms. Cardiorespiratory gating enabled an average 45% reduction in target expansion due to cardiac motion. Comparison of CardioKit respiratory motion traces with respiratory 4D-CT yielded differences of 1.1-1.4 mm in total motion amplitude and 0.97-0.99 Pearson correlation factor. These findings are valuable for establishing a workflow for cardiorespiratory-gated STAR treatment with CardioKit.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Validation of Cardiorespiratory-Gated Proton Beam Delivery for Stereotactic Arrhythmia Radioablation.Journal of cardiovascular electrophysiology · 2026
- Development of a prospective clinical workflow for proton beam therapy of ventricular tachycardiaStrahlentherapie und Onkologie · 2026
- STAR in a Second: Ultra-High-Dose-Rate Spread-Out Bragg Peak Proton Therapy for Breath-Hold Stereotactic Arrhythmia Radioablation.Advances in radiation oncology · 2026
- Integration of an echocardiographic motion-management system into photon-based stereotactic arrhythmia radioablation planning.Physics and imaging in radiation oncology · 2026
- Stereotactic Body Radiotherapy for Cardiac Arrhythmias: Current Evidence and Clinical Boundaries.Trends in cardiovascular medicine · 2026
- Variability and Main Determinants of Planning Target Volume During Cardiac Stereotactic Radioablation for Refractory Ventricular Arrhythmias: A Systematic Review and Pooled AnalysisJournal of Cardiovascular Electrophysiology · 2026