Relevance of light conditions for clinical in situ Raman spectroscopy.
Sven Richter, Sotirios Kalousios, Roberta Galli, Jonathan Ziegler, Jakob Dremel, Achim Temme, Katrin Kirsche, Ilker Y Eyüpoglu,
Department of Neurosurgery, Faculty of Medicine and University Hospital Carl Gustav Carus, TUD Dresden University of Technology, Dresden, Germany. · Medical Physics and Biomedical Engineering, Faculty of Medicine, TUD Dresden University of Technology, Dresden, Germany.
Key points
From the paper's conclusion- For clinical Raman spectroscopy, it is of utmost importance to be aware of optical disturbances. Intense illumination needs to be avoided but (spectrally structured) disturbances can be corrected by data preprocessing. Rather than external illumination, strong tissue autofluorescence might lead to impaired Raman spectral quality, making tumor tissue recognition difficult.
Abstract
BACKGROUND: Clinical Raman spectroscopy is a promising optical technology for intraoperative brain tumor delineation and diagnosis. As it exploits weak signals, illumination of the operating room (OR) and OR devices might impact the acquired spectra. The aim of this study was to investigate which optical artifacts can be tolerated during the intraoperative analysis of brain tumors in situ while still allowing the extraction of meaningful Raman spectra. METHODS: Tissue samples of rat brain and human brain tumors were investigated in an experimental OR, and background signals and Raman spectroscopic datasets were acquired using a fiber-based system. The measurements were performed under various lighting conditions and OR equipment. RESULTS: Ceiling light intensity increased the background, which could be prevented by manual coverage. Hence, the signal-to-noise-ratio of Raman spectra was not altered by ceiling lights. The intensity of the raw spectra of brain tissue and glioma was determined mainly by tissue autofluorescence and not by OR illumination, but it substantially affected the raw spectra of meningioma having low autofluorescence. Operating (OP) lights lead to saturation of the Raman camera. The exoscope, displays, and optical neuronavigation systems produced spectrally structured artifacts. Correct background subtraction was sufficient for artifact correction and obtaining valid Raman spectra. CONCLUSIONS: For clinical Raman spectroscopy, it is of utmost importance to be aware of optical disturbances. Intense illumination needs to be avoided but (spectrally structured) disturbances can be corrected by data preprocessing. Rather than external illumination, strong tissue autofluorescence might lead to impaired Raman spectral quality, making tumor tissue recognition difficult.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Robustness of Raman Spectroscopy‐Based Brain Tissue Classification Under Simulated Measurement DistortionsJournal of Raman Spectroscopy · 2026
- Shining a light on brain tumors: a review of Raman spectroscopy in the operating room.Journal of biomedical optics · 2026
- Advances in label-free Raman spectroscopic techniques for intraoperative tumor margin delineation: a comprehensive reviewApplied Spectroscopy Reviews · 2026
- Feasibility study on the application of Raman spectroscopy in the diagnosis of glioma.Translational cancer research · 2026
- Raman Spectroscopy for Glioma Diagnosis Toward Precision Neuro-Oncology: A Systematic Review.Critical reviews in oncology/hematology · 2026
- The Intraoperative Utility of Raman Spectroscopy for Neurosurgical OncologyCancers · 2025