Validation of a long-read 16S rRNA-gene sequencing approach for analysis of clinical samples and bacterial identification in a routine clinical laboratory.
Otto Van de Gaer, Reinoud Cartuyvels, Timo Froyen, Petra Hilkens, Koen Magerman, Steven Martens, Britt Van Meensel, Jozef Dingemans
Department of Laboratory Medicine, Jessa Hospital, Hasselt, Belgium.
Key points
From the paper's conclusion- An estimated limit of detection of 260 and 747 CFU/ml was found for respectively Staphylococcus aureus and Pseudomonas aeruginosa in synovial fluid. R2 analysis ranged from 0.9778 to 0.9986 depending on sample matrix. Correct genus level identification of strains was >98% and correct species level identification was >85%. 25/26 clinical culture-positive samples tested positive with 16S sequencing. Additional clinically relevant pathogens were detected in 5/28 culture-negative and 7/26 culture-positive samples.
Abstract
BACKGROUND: Currently, microbiological diagnosis relies heavily on classical culture-based techniques, which have notable limitations. These methods often struggle with detecting non-culturable or slow-growing microorganisms and can yield negative results in patients who have already received antimicrobial treatment before sample collection. 16S rRNA gene sequencing offers an alternative to culture-based methods, enabling bacterial identification without the need for cultivation. This study presents the validation and clinical implementation of a 16S rRNA-gene Nanopore-based sequencing workflow in a routine hospital. METHODS: Estimated limit of detection and linearity were investigated using spiking experiments with a dilution series of ATCC strains and commercial quality control material. Accuracy and including taxonomic resolution was evaluated for 37 strains obtained from external quality control programs and 30 clinical from a clinical collection. Utility was investigated by comparing 16S rRNA gene Nanopore-based sequencing with culture results and other clinical findings for 54 prospectively collected clinical samples. RESULTS: An estimated limit of detection of 260 and 747 CFU/ml was found for respectively Staphylococcus aureus and Pseudomonas aeruginosa in synovial fluid. R2 analysis ranged from 0.9778 to 0.9986 depending on sample matrix. Correct genus level identification of strains was >98% and correct species level identification was >85%. 25/26 clinical culture-positive samples tested positive with 16S sequencing. Additional clinically relevant pathogens were detected in 5/28 culture-negative and 7/26 culture-positive samples. DISCUSSION: This study demonstrates that the implementation of 16S rRNA gene sequencing is feasible in a routine clinical laboratory setting and provides additional utility for the identification of bacterial isolates and direct detection on clinical samples, increasing recovery of non-culturable isolates.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Clinical Accuracy and Operational Evaluation of a Novel Nanopore Based 16S Diagnostic Pathogen Detection Workflow.The Journal of infection · 2026
- Bacterial identification in clinical samples - comparing Nanopore and Illumina sequencing of the 16S rRNA and rpoB genes.BMC infectious diseases · 2026
- A streamlined workflow for nanopore-based full-length 16S rRNA gene profiling of low-biomass clinical specimens: application to the seminal microbiomeScientific Reports · 2026
- Rapid diagnosis of common, undetected, and uncultivable bloodstream infections from positive blood cultures using Oxford Nanopore sequencing: a metagenomic pipeline analysisThe Lancet Microbe · 2026
- A comparison of two culture and three sequencing methods for detection of the urobiomebioRxiv · 2026
- Evaluation of full-length 16S rRNA amplicon sequencing using Oxford Nanopore Technologies for diversity surveys of understudied microbiomes.Access microbiology · 2026