Surface engineering and coatings for metallic orthopaedic implants: an integrative review of biological performance and clinical translation
Ricardo Luiz Perez Teixeira
Key points
From the paper's conclusion- Coatings modify metallic orthopaedic implant interfaces, but their clinical value depends on biological safety and function during service.
Abstract
Coatings modify metallic orthopaedic implant interfaces, but their clinical value depends on biological safety and function during service. This integrative review examines bone-facing arthroplasty interfaces and internal fracture fixation. It links infection control and bone integration to corrosion, wear and coating integrity across titanium, stainless steel, cobalt–chromium and magnesium substrates. The original corpus comprised supplied bibliographic exports and targeted additions, rather than a comprehensive, reproducible search. After exclusion of two unavailable full texts, the review retained 27 primary reports. The sole author verified animal judgements; laboratory appraisal used descriptive criteria. Independent dual appraisal was not performed. Additional published clinical, retrieval, and mechanistic literature broadens the synthesis beyond the retained appraisal denominator. Experimental studies support biological activity under specific conditions. Randomised trials of antibiotic-loaded hydrogel coatings provide evidence for early infection prevention in defined settings. Conversely, long-term knee comparisons and registry analyses do not demonstrate a general clinical advantage from coating alone. Magnesium requires time-matched assessment of degradation, residual support, and healing. Heterogeneity in composition, implant role, exposure, and outcomes prevents a defensible ranking of coating families. A candidate antimicrobial exposure range must preserve host-cell function and integration on the same finished surface. Its persistence after sterilisation, ageing and mechanical damage requires separate evaluation. The review contributes a framework for assessing concurrent function and device-specific durability. Limited ascertainment and single-reviewer appraisal restrict the conclusions; laboratory activity alone cannot establish clinical benefit.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Comparative effectiveness of antimicrobial implant surface coatings in preventing orthopaedic implant-associated infections: a network meta-analysisArchives of Orthopaedic and Trauma Surgery · 2026
- From Conventional Biomaterials to Smart Bioactive Interfaces: Surface Engineering Strategies for Next-Generation Orthopedic ImplantsApplied Sciences · 2026
- Engineered surfaces for biomedical implants: advances in coatings, materials, and techniquesProgress in Biomedical Engineering · 2026
- Bioactive coatings for biomedical engineering application: A review of mechanical integrity, biocompatibility, and future smart technologies.Science progress · 2026
- Progress and Current Status of Biodegradable Metallic Implants: Mini-ReviewCorrosion and Materials Degradation · 2026
- Additive manufacturing of metallic implants: a comprehensive review of material selection, porous architecture, and hybrid surface engineeringFrontiers in Mechanical Engineering · 2026