Ventilator brand, not ventilation mode, determines electricity consumption and estimated carbon emissions: a bench study of nine intensive care ventilators.
Semahattin Serkan Sezer, Nimet Şenoğlu
Department of Intensive Care, Çiğli Training and Research Hospital, İzmir Bakırçay University, İzmir, Türkiye.
Key points
From the paper's conclusion- The findings support requiring measured power to be declared and evaluated as a standardised criterion in ventilator tenders; because all devices were approximately ten years old, they do not identify which currently marketed platform is preferable.
Abstract
BACKGROUND: Health care generates 4%-5% of global greenhouse gas emissions, and the intensive care unit (ICU) is among its most energy-intensive environments, where energy rather than consumables dominates the carbon footprint. Ventilators run continuously, yet their electricity consumption has rarely been measured directly and never compared between brands. METHODS: Active power was measured with a single plug-in digital meter on three devices from each of three brands, in every available ventilation mode plus standby (90 measurements). The device was the independent experimental unit, giving an effective sample of nine ventilators. All used integral internal gas delivery, with a shared circuit and test balloon under standardised settings. Power was read every 60 s over 10 min per mode and averaged, then modelled by linear mixed effects (brand and harmonised mode as fixed effects, random intercept for device within brand, brand-specific residual variance). The brand effect was corroborated by parametric bootstrap, the mode effect by equivalence testing; a brand-by-mode interaction was fitted. Emissions were projected across several utilisation and grid scenarios. RESULTS: Estimated marginal means were 64.9 W (Brand A), 95.4 W (Brand B) and 56.3 W (Brand C). Brand exerted a strong effect (bootstrap p = 0.003 ); mode showed no detectable effect under the tested conditions (bootstrap p = 0.51 ), every mode was equivalent to the reference at a pre-defined ± 10 W margin, and no brand-by-mode interaction was detected ( p = 0.27 , limited power). Brand B exceeded Brand C by 39.1 W (95% CI 17.0-61.2): a projected 161 kg CO 2 -equivalent (95% CI 70-252) per ventilator annually at the Turkish grid factor under continuous operation, and 17 to 161 kg across scenarios. Standby reached 92.6% of active consumption in one brand; the active-minus-standby increment (4.8, 28.7 and 15.2 W) placed most of the between-brand difference in the idle baseline rather than in ventilation. CONCLUSIONS: Brand, not ventilation mode, dominated electricity consumption among the devices tested. The findings support requiring measured power to be declared and evaluated as a standardised criterion in ventilator tenders; because all devices were approximately ten years old, they do not identify which currently marketed platform is preferable.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Association Between Mechanical Power and Ventilator-Associated Conditions in Patients Requiring Prolonged Mechanical Ventilation: A Multi-Database Retrospective Cohort Study.Journal of clinical medicine · 2026
- Decarbonising Respiratory Care: The Impact of a Low‐Carbon Salbutamol Pressurised Metered‐Dose InhalersAllergy · 2025
- The role of timely assessment of appropriate care in reducing greenhouse gas emissions in intensive care medicine.Annals of intensive care · 2026
- Flow-controlled ventilation versus pressure-controlled ventilation in moderate to severe ARDS patients: a randomized crossover physiological studyIntensive Care Medicine Experimental · 2025
- Electricity Grid Carbon Intensity-Not Imaging Utilization-as the Primary Source of Variation in Greenhouse Gas Emissions for MRI and CT.AJR. American journal of roentgenology · 2026
- Bench evaluation of six non-invasive ventilation home ventilators: comparison with an ICU ventilator and unsupervised clusteringBMJ Open Respiratory Research · 2025