Case report: When quality control goes wrong — an incidental finding of Hb E compound heterozygosity with Filipino β0-thalassemia deletion in a commercial EQA material mistaken as Hb E homozygote

Juan Tang, Liang Liang, Yun Yu, Lihong Zheng, Xiong Huang, Youqiong Li

Case report

Key points

From the paper's conclusion
  • After Nanopore-based TGS rechecking, the Filipino deletion was detected, and cell line trace-back confirmed compound heterozygosity for Hb E and the Filipino β 0 -thalassemia deletion.

Abstract

Objective Hb E homozygosity and compound heterozygosity for Hb E and β 0 -thalassemia show highly similar profiles on hemoglobin analysis, creating a diagnostic pitfall. We aimed to resolve an incidental genotype discrepancy in an external quality assessment (EQA) sample originally labeled as Hb E homozygote. Methods Routine thalassemia genetic testing covered 26 mutations commonly found in the Chinese population. Multiplex ligation-dependent probe amplification (MLPA) was used to screen for potential large deletions or duplications. Third-generation sequencing (TGS) was performed to define the exact location of the large deletion, and the breakpoints it identified were subsequently validated by Gap-PCR and Sanger sequencing. Results The sample showed Hb A 0%, Hb A 2 0%, Hb F 43.6%, Hb E 56.4%, Hb 8.3 g/dL, and MCV 64.5 fL. PCR-reverse dot blot hybridization indicated Hb E homozygosity. MLPA revealed heterozygous deletions across multiple probes, consistent with the Filipino deletion. TGS identified a 118,475 bp deletion at chr11:5112883-5231358del. Gap-PCR and Sanger sequencing confirmed a 118,475 bp deletion. The supplier initially disputed these findings and reported Hb E homozygosity without deletion by NGS. After Nanopore-based TGS rechecking, the Filipino deletion was detected, and cell line trace-back confirmed compound heterozygosity for Hb E and the Filipino β 0 -thalassemia deletion. Conclusion EQA samples with complex or overlapping hemoglobin phenotypes require comprehensive molecular validation before distribution. Reliance on a single assay or manufacturer-provided assignment can lead to erroneous EQA evaluation.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers