Preliminary laboratory analytical comparison of quantitative at-home urinary hormone monitoring systems using artificial urine.
Thomas P Bouchard, Paul J Yong, Patricia K Doyle-Baker
Department of Family Medicine, University of Calgary, Calgary, AB, Canada. · Department of Obstetrics & Gynecology, University of British Columbia, Vancouver, BC, Canada.
Key points
From the paper's conclusion- This preliminary laboratory comparison aimed to provide an analytical framework to compare multiple commercially available at-home urine hormone monitoring devices across key performance metrics under controlled artificial urine conditions.
Abstract
BACKGROUND: Quantitative at-home urinary hormone monitoring systems are increasingly used for reproductive health tracking, but their analytical behaviour requires systematic characterization. This preliminary laboratory comparison aimed to provide an analytical framework to compare multiple commercially available at-home urine hormone monitoring devices across key performance metrics under controlled artificial urine conditions. METHODS: Laboratory experiments were informed by Clinical and Laboratory Standards Institute's (CLSI) validation principles. Four urine hormone monitoring devices were evaluated using a pre-registered analytical framework for luteinizing hormone (LH), follicle-stimulating hormone (FSH), estrone-3-glucuronide (E3G), and pregnanediol glucuronide (PDG). Metrics included coefficients of variation (CV, for precision), relative deviation (RD, for accuracy), linearity, limit of blank (LoB), limit of detection (LoD) for each device, using artificial urine samples with standard hormone concentrations. Cross-reactivity was tested with other hormones including human chorionic gonadotropin (hCG), cortisol, thyroid stimulating hormone (TSH) and progesterone subtypes. Device-specific evaluations, including the impact of ambient light and various camera systems, were also conducted for relevant devices. RESULTS: There were device- and analyte-specific differences in precision, accuracy, linearity and detection limits. Precision and accuracy ranged from CV/RD values near the 15% acceptability threshold to levels exceeding 50% at some concentrations. External lighting conditions affected camera-based strips at low hormone concentrations. False positive results for LH can occur in the presence of 100 mIU/mL hCG, which could lead to misinterpretation in the context of pregnancy. One device that measures the LH-beta subunit avoided this cross-reactivity. CONCLUSIONS: A preliminary analytical framework was applied to evaluate the performance of quantitative urinary hormone devices for at-home use. This framework can be used in future research and development of menstrual health monitoring technologies. Longitudinal clinical studies using human urine samples and appropriate ovulation reference standards are required before conclusions can be drawn regarding ovulation detection, fertile-window estimation or menstrual-phase classification.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Quantitative urinary hormone biomarkers for ovulation detection and timing: a comparison with transvaginal ultrasonography and serum hormonesFrontiers in Endocrinology · 2026
- Utility and practical application of at-home daily urine hormone monitoring to support health, nutrition, and exercise: illustrative case-based perspectives across the female lifespan.Postgraduate medicine · 2026
- A Novel Approach to Characterization of the Female Hormone Profile: Feasibility and Research Application of Daily At-Home Urine Hormone Monitoring.Medicine and science in sports and exercise · 2026
- HueFertility Ovulation: A Smartphone-Enabled Quantification System for Urinary Luteinizing Hormone in a Lateral Flow Assay.Biosensors · 2026
- Timing luteal urinary pregnanediol glucuronide assessment using urinary luteinizing hormone: agreement with ultrasound-determined ovulation.Frontiers in endocrinology · 2026
- Multiplex ovarian hormone monitoring in sweat using a dual-aptamer wearable sensor.Device · 2026