Camera-based HbA1c estimation using remote photoplethysmography: development, independent evaluation, and baseline comparator analysis.
Hairil Rizal Abdullah, Si Ying Tan, Isuru Kalhara, Shiqi Wang, Vera Jia Qi Lee, Roderica Rui Ge Ng, Yong Mong Bee
Department of Anaesthesiology, Singapore General Hospital, Singapore, Singapore. · Nervotec, Singapore, Singapore.
Key points
From the paper's conclusion- We tested whether facial rPPG color time series carry information about laboratory HbA1c beyond simple comparators.
Abstract
Type 2 diabetes mellitus is highly prevalent and frequently undiagnosed, yet HbA1c testing depends on blood sampling and laboratory infrastructure, limiting its use for low-friction dysglycemia screening. Remote photoplethysmography (rPPG) records facial color signals with a standard camera and has been proposed as a digital biomarker platform for cardiometabolic risk. We tested whether facial rPPG color time series carry information about laboratory HbA1c beyond simple comparators. In a prospective single-center study at Singapore General Hospital, 254 participants had both a usable facial recording and a laboratory HbA1c. Each recording was reduced to the mean red, green and blue values of six facial regions over 8.53 s. NervoHbA1c, a 4.27-million-parameter convolutional model with demographic inputs, was evaluated by pooled 5-fold cross-validation repeated three times, leave-one-cohort-out evaluation, a learning curve and testing on two independent smartphone cohorts, against mean, median and demographics-only comparators fitted within each training split. Within the development cohort, 16 models spanning 0-4.27 million fitted parameters were compared, including frozen-encoder linear read-outs and conventional signal-processing features. Pooled over 254 participants, NervoHbA1c had a mean absolute error (MAE) of 1.081% (95% CI 0.973-1.193), against 1.230% for the median baseline, 1.237% for demographics-only and 1.254% for the mean baseline (Nadeau-Bengio p = 0.008-0.042), and its predictions correlated with laboratory HbA1c (r = 0.505). It had the lowest error in every leave-one-cohort-out split, and its error fell steadily as the training set grew. In the development cohort, error decreased with model capacity: every lower-capacity model had higher error than the full model (MAE 0.732% against 0.844% for the mean baseline). A signal-only model was modestly correlated with HbA1c (r = 0.368). The published checkpoint, trained on 53 webcam participants, did not transfer to the smartphone cohorts (SGH-6 MAE 1.478% against 1.205% for demographics-only), a gap attributable to the change of acquisition device (webcam against smartphone AUROC 1.000). Short facial rPPG recordings therefore carry modest but recoverable information about HbA1c beyond demographics when the training data match the acquisition device. Participant-level evaluation, cross-cohort testing and simple comparators are needed before camera-based digital biomarkers advance toward clinical use, and a larger full-precision cohort is planned.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Photoplethysmography-based glucose estimation with feature selectionbioRxiv · 2026
- AI-powered cardiometabolic assessment from a single facial videoEuropean Heart Journal · 2026
- Uncertainty-aware fourier-encoded attentive Bi-GRU for non-invasive blood glucose estimation from multimodal PPG and GSR signals.Frontiers in medical technology · 2026
- Adaptive digital filtering and TreeSHAP feature selection for non-invasive diabetes classification from photoplethysmography signalsDiscover Artificial Intelligence · 2026
- Minute-Scale Sleep-Onset HRV as a Candidate Risk-Stratification Biomarker for Diabetic Retinopathy: An Exploratory Pilot Study.Diagnostics (Basel, Switzerland) · 2026
- Conditional Diffusion-Augmented Cascaded Multi-View Attention BiLSTM for Non-Invasive Blood Glucose Estimation from Photoplethysmography.Bioengineering (Basel, Switzerland) · 2026