Balancing innovation and safety in percutaneous treatment of coronary artery calcification: the evolving role of cutting balloon technology.

Pierre Chenard, Claudiu Ungureanu, Giuseppe Colletti, Adrien Jossart, Daciana Popa, Agostino Spano, Gregor Leibundgut, Marouane Boukhris

Department of Cardiology, CHU Dupuytren, Limoges, France. · Department of Cardiology, Hôpital de Jolimont, Hélora Network, La Louvrière, Belgium.

Review

Key points

From the paper's conclusion
  • Current evidence does not justify routine high-pressure iterative CB use beyond device labelling or universal IDE-guided treatment targets.

Abstract

BACKGROUND: Cutting balloon (CB) angioplasty is one of several approaches to lesion preparation in calcified coronary artery disease. Its comparative efficacy, safety, and optimal use remain incompletely defined. AIMS: To critically appraise clinical studies and emerging technical concepts concerning CB angioplasty, with particular attention to evidence strength, patient selection, and safety. METHODS: This narrative review discusses selected randomised trials, observational studies, mechanistic investigations, and emerging reports concerning CB-based lesion preparation. FINDINGS: Randomised studies have reported improvements in selected acute imaging endpoints, but findings differ across populations and treatment strategies. Observational studies suggest progressive lumen gain during sequential CB inflations; however, these observations do not establish an optimal inflation protocol. Evidence for comparative long-term clinical benefit and rare device-related complications remains limited. Reported procedural cost differences should not be equated with generalisable cost-effectiveness. Inflation above the device's rated burst pressure constitutes off-label use and falls outside the manufacturer's recommended pressure limits. The Index of Device Efficiency (IDE) remains an exploratory metric rather than a validated treatment target. CONCLUSION: CB may be considered among complementary calcium-modification options in appropriately selected lesions. Device choice should reflect anatomy, crossability, imaging findings, and patient risk. Current evidence does not justify routine high-pressure iterative CB use beyond device labelling or universal IDE-guided treatment targets.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.

Related papers