Balancing innovation and safety in percutaneous treatment of coronary artery calcification: the evolving role of cutting balloon technology.
Pierre Chenard, Claudiu Ungureanu, Giuseppe Colletti, Adrien Jossart, Daciana Popa, Agostino Spano, Gregor Leibundgut, Marouane Boukhris
Department of Cardiology, CHU Dupuytren, Limoges, France. · Department of Cardiology, Hôpital de Jolimont, Hélora Network, La Louvrière, Belgium.
Key points
From the paper's conclusion- Current evidence does not justify routine high-pressure iterative CB use beyond device labelling or universal IDE-guided treatment targets.
Abstract
BACKGROUND: Cutting balloon (CB) angioplasty is one of several approaches to lesion preparation in calcified coronary artery disease. Its comparative efficacy, safety, and optimal use remain incompletely defined. AIMS: To critically appraise clinical studies and emerging technical concepts concerning CB angioplasty, with particular attention to evidence strength, patient selection, and safety. METHODS: This narrative review discusses selected randomised trials, observational studies, mechanistic investigations, and emerging reports concerning CB-based lesion preparation. FINDINGS: Randomised studies have reported improvements in selected acute imaging endpoints, but findings differ across populations and treatment strategies. Observational studies suggest progressive lumen gain during sequential CB inflations; however, these observations do not establish an optimal inflation protocol. Evidence for comparative long-term clinical benefit and rare device-related complications remains limited. Reported procedural cost differences should not be equated with generalisable cost-effectiveness. Inflation above the device's rated burst pressure constitutes off-label use and falls outside the manufacturer's recommended pressure limits. The Index of Device Efficiency (IDE) remains an exploratory metric rather than a validated treatment target. CONCLUSION: CB may be considered among complementary calcium-modification options in appropriately selected lesions. Device choice should reflect anatomy, crossability, imaging findings, and patient risk. Current evidence does not justify routine high-pressure iterative CB use beyond device labelling or universal IDE-guided treatment targets.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Greater stent expansion at the maximally calcified segment with cutting balloon versus noncompliant balloon preparation: the randomized PREDICTION trial.Cardiovascular intervention and therapeutics · 2026
- Balancing innovation and safety in percutaneous treatment of coronary artery calcification.Frontiers in cardiovascular medicine · 2026
- Calcium-Modification Techniques in Calcified Coronary Artery Disease: A Systematic Review of Comparative Evidence on Procedural Success, Complications and Clinical Outcomes.Journal of clinical medicine · 2026
- Stonebreaker: Current Approaches to Percutaneous Coronary Intervention of Calcified Coronary Lesions.Medicina (Kaunas, Lithuania) · 2026
- Optimizing Drug-Coated Balloon Outcomes: The Evolving Role of Vessel Preparation in Coronary and Peripheral Artery Interventions.Life (Basel, Switzerland) · 2026