Frontiers in bioengineering and biotechnology

FRESH-bioprinted vascular constructs: a quantitatively characterized workflow for reproducible device-oriented applications.

Franca Scocozza, Elena Delgrosso, Michele Conti

Department of Civil Engineering and Architecture, University of Pavia, Pavia, Italy.

Key points

From the paper's conclusion
  • Here, we establish a quantitatively characterized FRESH-bioprinted workflow that enables reproducible fabrication and standardized characterization of vessel-like constructs for device-oriented testing.

Abstract

The lack of standardized and reproducible in vitro vascular models remains a major obstacle to the quantitative evaluation of endovascular devices and vessel-device interactions. While recent bioprinted vascular constructs have increased biological complexity, reproducibility, shape fidelity, and mechanical consistency remain insufficiently characterized. Here, we establish a quantitatively characterized FRESH-bioprinted workflow that enables reproducible fabrication and standardized characterization of vessel-like constructs for device-oriented testing. Bioink formulation, printing parameters, and post-processing conditions were systematically optimized. At the same time, construct geometry and mechanical reproducibility were quantitatively evaluated using micro-computed tomography (micro-CT), confined compression, and ring tensile testing. Sterile fabrication, preliminary cellularization, controlled perfusion, and intraluminal stent-like ring deployment were further explored to assess procedural compatibility within a device-oriented vascular testing framework. The optimized workflow enabled reproducible fabrication of vessel-like constructs across multiple geometrical scales, with close agreement between designed and measured dimensions and limited batch-to-batch variability. Micro-CT analysis revealed controlled post-processing-induced geometrical changes, including mild shrinkage and lumen reduction, while mechanical testing demonstrated reproducible axial and circumferential responses across independent fabrication batches. Intraluminal deployment of a radiopaque polycaprolactone ring induced local vessel expansion of approximately 8% while preserving lumen patency and structural integrity under low-pressure perfusion conditions. Preliminary biological assays confirmed high cell viability for up to 14 days and initial endothelial adhesion, supporting compatibility with sterile cell culture procedures and providing a robust foundation for future biologically advanced vascular models. By integrating biofabrication, quantitative imaging, mechanical characterization, and preliminary biological validation into a single workflow, this work establishes a quantitatively characterized and reproducible FRESH-bioprinted vascular platform that bridges the gap between rigid vascular phantoms and biologically advanced tissue-engineered systems. The proposed platform provides a robust experimental foundation for future process standardization, device-oriented vascular testing, and the progressive development of standardized experimental models for endovascular research.

Read the paper

Discussion

Science Skim's AI key points, plus readers' public thoughts

Readers' thoughts

No reader thoughts yet.