Anin vitro, proof of concept study utilizing a latex model to simulate assessment of urethral pressure profile to quantify tissue compliance and stricture detection.
Diboro L Kanabolo, Noah Roselli, Emily Ji, Yashwanth Nanda Kumar, Neelesh A Patankar, Ziho Lee
Department of Urology, Northwestern Memorial Hospital, Chicago, IL, USA. · Department of Engineering Sciences and Applied Mathematics, Northwestern University, Chicago, IL, USA.
Key points
From the paper's conclusion- This proof-of-concept model demonstrates that pressure profiling can distinguish healthy from strictured urethras and supports compliance as a novel diagnostic metric.
Abstract
BACKGROUND: Retrograde urethrogram (RUG) remains the gold standard for diagnosing urethral stricture disease, however no widely available diagnostic modality currently quantifies stricture characteristics using cross-sectional area or tissue compliance. We hypothesized that pressure-derived metrics could improve diagnostic accuracy and guide treatment. We aimed to develop an in vitro latex urethra model to simulate "healthy" and "strictured" states and directly measure luminal pressure as an initial step toward compliance assessment. METHODS: Ten 20-cm latex tubes were studied; five contained 1.5-cm simulated strictures. All tubes were clamped at both ends to prevent leakage. Strictures were created by clamping across an 8Fr bougie at the midpoint, while healthy models were left unclamped at the 8Fr bougie location. Tubes were pressurized for 10 s to simulate RUG conditions. Pressure was measured using a 7Fr air-charged dual-sensor urodynamic catheter positioned such that the midpoint stricture was located 2 cm distal to the first pressure sensor. Saline was infused at 20 mL/min through the catheter port positioned proximal to the stricture, and maximum pressure (Pmax, cmH2O) was selectively recorded only from the first proximal sensor to reflect upstream resistance generated by the narrowing. Time to maximum pressure (Tmax, seconds) was also measured. Compliance was calculated (ΔV/ΔP). Comparisons were performed using an unpaired Student's t-test. RESULTS: Ten pressure-time curves were generated. Mean (Standard Deviation) Pmax was significantly higher in strictured models compared to healthy controls (160.8 [9.4] vs. 29.8 [16.4] cmH2O, p < 0.001). Tmax was similar between groups (9.2 [1.1] vs. 9.8 [0.4] s, p = 0.289). Compliance was lower in strictured (0.019 [0.003] cm³/cmH2O) vs. healthy tubes (0.150 [0.107] cm³/cmH2O, p = 0.025). CONCLUSIONS: This proof-of-concept model demonstrates that pressure profiling can distinguish healthy from strictured urethras and supports compliance as a novel diagnostic metric.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Development of a cost-effective bladder-prostate phantom model to simulate the effects of minimally invasive surgical treatments (MIST) for benign prostatic hyperplasia (BPH) and its validation by pressure-flow study.Biomedical physics & engineering express · 2026
- Diagnosis of Urethral Resistance in Men: 70 Years of History.Neurourology and urodynamics · 2026
- UROREC: A cystometry and electromyography recording system for analyzing urinary incontinence parameters in rat models.PloS one · 2026
- A high-resolution wireless sensing platform for the characterization of urodynamic events in murine models.Biomedical physics & engineering express · 2026
- Transrectal Biplane Ultrasound Combined with Cotton Swab Marking of the External Urethral Orifice for Evaluating the Postoperative Position of Transobturator Mid-Urethral Sling: A Study on Standardized Ultrasound Measurement and Inter-Observer Reliability.Diagnostics (Basel, Switzerland) · 2026
- Feasibility Study on Non-Invasive Methods for Bladder Biomechanics During Voiding Using Uro-Dynamic MRI.Bioengineering (Basel, Switzerland) · 2026