Combined accelerometry and writing analysis for quantifying tremor during focused ultrasound thalamotomy
Andrew E Toader, Nemanja Useinovic, Beck Shafie, Matthew C. Henn, M. Patricia Joyce, Haley D. Smith, Lee E. Neilson, Delaram Safarpour,
Oregon Health & Science University
Key points
From the paper's conclusion- Using these methods, the mean ± standard error of the improvements in tremor after MRgFUS thalamotomy was 63.6% ± 7% by accelerometer analysis, 72.4% ± 9.9% by written spiral analysis, 68% ± 6.7% by combining the accelerometer and written spiral analyses, and 33.6% ± 3.1% by TETRAS analysis.
Abstract
OBJECTIVE: MRI-guided focused ultrasound (MRgFUS) thalamotomy is used to treat tremors. Identifying the correct area to ablate relies on subjective physician assessment of tremor improvement during treatment. To address this limitation, objective quantification of tremor severity during MRgFUS is needed. The aim of this study was to develop and evaluate an objective method for quantifying tremor progression during MRgFUS using an MRI-compatible accelerometer and tablet. METHODS: Forty patients (30 male, mean age 74.5 years) who underwent MRgFUS thalamotomy to treat tremor were included in this analysis. All patients were evaluated during the procedure using analysis of drawn Archimedean spirals and a pen-mounted accelerometer. The severity of the tremor was determined by analysis of the rhythmic oscillatory patterns present in the drawings and accelerometer recordings after each sonication delivered. The patient's drawings were evaluated by two movement disorder neurologists using the drawing subsection of The Essential Tremor Rating Assessment Scale (TETRAS), and subsequently compared with the TETRAS score. RESULTS: Using these methods, the mean ± standard error of the improvements in tremor after MRgFUS thalamotomy was 63.6% ± 7% by accelerometer analysis, 72.4% ± 9.9% by written spiral analysis, 68% ± 6.7% by combining the accelerometer and written spiral analyses, and 33.6% ± 3.1% by TETRAS analysis. When stratifying based on temperature, improvement was 22.8% ± 4.4% at < 50°C, 48.9% ± 5.5% at 50°C-53°C, and 70.8% ± 2.9% at > 53°C. CONCLUSIONS: The measurements from the accelerometer and written spiral analysis were comparable with expert analysis of TETRAS scores, but the accelerometer had greater sensitivity for subtle improvement in tremor. This system of analysis provides an objective and instantaneous measure of tremor improvement during MRgFUS, potentially making the procedure safer and more efficient.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Less is more: comparison of automated spiral-drawing metrics with clinician-rated tremor scores after MRgFUS thalamotomy.Frontiers in neurology · 2026
- Focused Ultrasound for Essential Tremor in Denmark: Results After 1 Year of Follow‐UpMovement Disorders · 2026
- Tremor Morphology Toolkit: Waveform Modulation-Based Tremor Phenotyping.medRxiv : the preprint server for health sciences · 2026
- Longitudinal Patient-Reported Outcomes in Essential Tremor and Tremor-Dominant Parkinson's Disease Treated with MR-Guided Focused Ultrasound Vim Thalamotomy: A Real-World Cohort Study.Brain sciences · 2026
- Unilateral magnetic resonance-guided focused ultrasound for medication-refractory essential tremor: 5-year continued access studyFrontiers in Neurology · 2025
- Interrater reliability of Tremor Characterization.medRxiv : the preprint server for health sciences · 2026