Development and Validation of Persian Speech Test Signal (PerSTS).
Sasan Etezadi, Nariman Rahbar, Seyyed Jalal Sameni, Farhad Torabinezhad, Behrooz Mahmoodi-Bakhtiari
Rehabilitation Research Center, Department of Audiology, School of Rehabilitation Sciences, Iran University of Medical Sciences, Tehran, Iran. · Rehabilitation Research Center, Department of Speech and Language Pathology, School of Rehabilitation Sciences, Iran University of Medical Sciences, Tehran, Iran.
Key points
From the paper's conclusion- The PerSTS mimics the acoustic and temporal features of natural Persian speech within International Electrotechnical Commission (IEC) standards, providing a standardized tool for hearing aid verification and audiology.
Abstract
BACKGROUND: Digital hearing aids require precise fitting to enable realistic speech processing. The international speech test signal (ISTS) provides a standard reference, yet does not fully capture the phonetic and prosodic features of Persian. This study aimed to create and acoustically verify a Persian version of the ISTS-termed the Persian speech test signal (PerSTS)-to enhance verification accuracy for Persian-speaking users. METHODS: The North Wind and the Sun (NWS) passage was translated into Persian with careful phoneme distribution and natural syntax. Six native female speakers recorded the passage at 44.1 kHz and 24-bit resolution. These recordings were segmented, enhanced, and stitched into 10-15 second sequences that adhered to pitch and spectral criteria. Sequences that closely matched the Byrne female long-term average speech spectrum (LTASS; ±1 dB) and were rated as most natural by linguists were assembled into a 60-second PerSTS. The final audio was calibrated to 65 dB sound pressure level (SPL) and analyzed for its spectral, temporal, and amplitude characteristics in Tehran, Iran (2025). RESULTS: The PerSTS exhibited spectral peaks between 200-400 Hz and 3-4 kHz, characteristic of female Persian speech. Its LTASS closely matches the Byrne LTASS and ISTS, with minor deviations below 500 Hz and above 8 kHz. The crest factor (16.7 dB) and dynamic range (27.1 dB) were consistent with those of natural speech. Experts rated its naturalness as 4.35±0.28 on a five-point scale. CONCLUSION: The PerSTS mimics the acoustic and temporal features of natural Persian speech within International Electrotechnical Commission (IEC) standards, providing a standardized tool for hearing aid verification and audiology.
Read the paper
Discussion
Science Skim's AI key points, plus readers' public thoughts
Readers' thoughts
No reader thoughts yet.
Related papers
- Stability and Test-Retest Reliability of the Individual Acoustic Reference for Within-Subject Analysis of Hearing Device-Processed Speech: The SpettroREM Method.Audiology research · 2026
- Development and evaluation of an Indian Language-Adjusted Gain model for hearing aid fitting in Malayalam and Kannada speakersScientific Reports · 2026
- An Analysis of Word-Level Intelligibility in EFL Learners Using ASRAhi Evran Üniversitesi Sosyal Bilimler Enstitüsü Dergisi · 2026
- Speech Signal Processing and Deep Neural Modeling for Pronunciation Evaluation in College English Teaching.Journal of visualized experiments : JoVE · 2026
- Cepstral Peak Prominence (CPP) Values in Dysphonic and Nondysphonic Hebrew Speakers.Journal of voice : official journal of the Voice Foundation · 2026
- Validation of FUDAN VOICE: Signal and Acoustic Feature Comparison with Computerized Speech Lab (CSL).Journal of voice : official journal of the Voice Foundation · 2026